INSTRUCTION MANUAL   

FOR 

Dust Monitor 

TYPE 

PFM-KCU01  (Transducer) 
PFM-M01PEX (Sensor) 
Z961/Z964 (Zener barrier) 

※ 

You,  an  excellent  operator,  please  read  the  manual  for  Model  thoroughly  and 
operate Model effectively for your factory and for yourself and thank you!! 

1-8-18 Norimatsu-Higashi,Yahatanishi-ku,Kitakyushu 807-0837 Japan 

Phone No.(8193)691-3731  Fax No.(8193)691-3735 
http://www.matsushima-m-tech.com 
E-mail [email protected] 

Summary of Contents for PFM-KCU01

Page 1: ... Maintenance 13 11 Trouble shooting 13 INSTRUCTION MANUAL FOR Dust Monitor TYPE PFM KCU01 Transducer PFM M01PEX Sensor Z961 Z964 Zener barrier You an excellent operator please read the manual for Model thoroughly and operate Model effectively for your factory and for yourself and thank you 1 8 18 Norimatsu Higashi Yahatanishi ku Kitakyushu 807 0837 Japan Phone No 8193 691 3731 Fax No 8193 691 3735...

Page 2: ... a correct wiring before applying power source Do not have a shock or strong impact to the equipment Failure to observe this CAUTION may result in equipment damage Be sure to connect necessary terminals grounding etc Remove all wiring to the equipment before doing electrical welding work near the equipment Do not forcedly bend or pull the lead wire also do not use unnecessarily long wire Tighten t...

Page 3: ...density specified Density indicated on specification is for design development of machinery and is publicized for you customers to use as a guideline in deciding the range When you customers decide the range you please find the range and set it in which alarm level is easy to set strictly from 0 100 signal and becomes pulse signal By doing so you can realize your management of exhaust drift If the...

Page 4: ...ity 2 95 RH Condition State of case sealing up 6 Mounting R1 Screw 1 Socket belonging 7 Wiring distance max 100m 3 Between Sensor and Zener barrier 5m Exclusive use cable Between Sensor and Zener barrier 0 4m Exclusive use cable 8 Length 4 Probe L1 300mm Isolation L2 58mm Permissible length 50 550mm 9 Mass Approx 1 3kg Zener barrier Z961 1 Approval Exib ⅡB 2 International Protection IP20 3 Rating ...

Page 5: ...execution of the withstand voltage test 4 Equipment specification Output unit 1 Power supply Please specify it AC110V 220V 15 10 50 60Hz 2 Power consumption 7VA 3 Output signal Analog signal DC4 20mA 1 max 23 3mA Load Max 500Ω Isolated Relay output signal 1 Alarm 1C AC250V 2A DC30V 2A 2 Upper limit 2C AC250V 2A DC30V 2A 4 Display Power Green ON 点灯 Alarm 1 Alarm Red ON 点灯 2 Upper limit Yellow ON 点灯...

Page 6: ...probe should be touched directly by hand without grounding while being energized 4 Please be sure to connect earth terminal 5 Please tighten the Bolt and Nut when you install it 6 Please make sure that the lead outlet is firmly closed so that the outside air does not invade inside the unit Where it is connected by piping construction such as conduit system please apply putty and so on to the lead ...

Page 7: ...m from plumbing pipe Important Wiring of power and signal should be independent please Cap is for transport protection and is not a raincoat Attention Exclusive cable is transmitting very weak signal Cable should be protected with by all means steel made electric wire pipe independently so that inductive noise form ambient area does not bother signal Securely close the lead outlet to prevent outsi...

Page 8: ... for Transducer attaching when you remove the attaching screw covers Please loosen these joining screws of the terminal block part with a Phillips driver You can see the terminal block of the Transducer when you remove the main block part from the terminal block part after having loosened these Joining Screws Joining Screws Mtg Hole Mtg Hole Mtg Hole Mtg Hole Fig 8 Joining screws of terminal and m...

Page 9: ...2 SPDT Ac250V 2A DC30V 2A Analog output signal DC4 20mA Load Max 500Ω Line connecting terminal Max 1 5mm2 Output unit Fig 10 Connection diagram Component part Packing Body Shield ring Rubber bush Clamping member Cap 2 Completion drawing 1 85 3 5 4 Packing 1 Please pass Cap Clamping member and Rubber bush through the cable Please peel off the cable covering from the verge by the length of 85mm 2 Pl...

Page 10: ...LED Light on as power on FAULT LED CPU on when abnormal L1 ON LED Light on over limit1set L2 ON LED Light on over limit1set 4 beams indicate LED detection density will indicate per 0 100 complying with setup for button use when changing beam of number value use when increasing number value MOD use when switching parameter set mode and measure mode Pushing after set change will memorize parameter a...

Page 11: ...utput due to its output linear Reference Talc Talc is non organic powder made from ore called Talc being crushed into gray white particle and smooth powder Talc with its chemical name hydrous magnesium silicate Mg3Si4O10 OH 2 containing main ingredient of SiO2 approx 60 MgO approx 30 and combined water of 4 88 As its physical nature Talc of true specific gravity 2 7 2 8 is a substance having lowes...

Page 12: ...n Customers expertise respectively Limit 1 2 and action chart of trouble alarm Power OFF to open COM NO Power OFF to close COM NO 20mA 4mA Time Range to set Limit set value ON Limit set value OFF Hysteresis COM NO close LED On COM NO open LED Off COM NO open LED Off Trouble Device normal Device normal COM NO close LED on COM NO open LED off COM NO open LED off Trouble alarm Limit output upper limi...

Page 13: ...ten Range on the left will light on 2 Push SEL button 5 times LED orange lit goes down in turn and INTEGRATION TIME LED will light 3 By using both button and button set Indication LED at 15 4 Push MOD button1 time return to measure window and start measure again at integral 15 second You operate other parameters with same procedure please In case setting other parameters continuously at 4 above in...

Page 14: ...rection sever itself and does not return Duct or plumbing pipe is short circuit from detection terminal due to terminal having sticking with material Terminal has dew condensation Electric charge does not generate Terminal is contacting with pipe inside GND Machinery is in trouble Big amount of material is flowing Attention 1 In case removing detection probe from machinery do it after loosening M5...