KAS/KAS-R series, -LRs/-LRs4 option
5-6
Command Descriptions
VSET command
Set the output voltage value.
Range
0 to the rated voltage value
Resolution
Refer to [5-7
Resolution of Command Setting and Return Values].
Unit
kV
Format ex.
VSET 12.34
⇒
Set 12.34kV to the output voltage.
Initial value
20
Query
In response to “VSET?”, it returns the set voltage value.
Remarks
・
Use the set value within the rated value even though the input range is up to the
rated voltage value multiplied by 102.5%.
・
This supports "Multi-command".
OUT command
Set OFF/ON of the output.
Range
OFF/ON
Format ex.
OUT ON
OUT OFF
Initial value
OUT OFF
Query
In response to “OUT?”, it returns OUT=OFF or OUT=ON.
Remarks
・
This supports "Multi-command".
REN command
Enable remote control.
Format
REN
Initial value
GTL
Remarks
・
When this command is received, all commands will become available.
・
Receiving this command will turn on the "REMOTE" LED.
・
This supports "Multi-command".
GTL command
Disable remote control.
Format
GTL
Initial value
GTL
Remarks
・
When this command is received, commands except for "REN", "STS", and output
measurement are ignored.
・
Receiving this command will turn off the "REMOTE" LED.
・
This supports "Multi-command".
UNIT command
Set the unit number.
Range
0 to 31
Format ex.
UNIT 0
UNIT 31
Initial value
UNIT 0
Query
In response to “UNIT?”, it returns the set unit number.
Remarks
・
This supports "Multi-command".