7 Operating with Digital Control
64
HARS series
7-8 Description of Original Command of Matsusada Precision
VSET Command
Sets the output voltage value.
Range
0 to Rated voltage value
Resolution
Refer to
[7-9 Range for Voltage Setting and Crrent Setting].
Unit
V
Format
example
VSET 23450
⇒
For V in unit, set 23.45kV.
Initial value
0
Query
In response to
“VSET?”, the set voltage value is returned in actual value.
Return example: VSET=23450
Remarks
・
Although the Input up to 102.5% of the rated voltage value is allowed, use it within
the rated value.
・
This command is compatible with "Multi-Command".
ISET Command
Sets the output current value.
Range
0 to Rated current value
Resolution
Refer to [7-9 Range for Voltage Setting and Crrent Setting].
Unit
Same as Unit displayed on the front panel.
Format
example
ISET 0.02345
⇒
For A in unit, set 23.45mA.
Initial value
0
Query
In response to
“ISET?”,
the set current value is retuned in actual value.
Return example: ISET=0.02345
Remarks
・
Although the Input up to 102.5% of the rated current value is allowed, use it
within the rated value.
・
This command is compatible with “Multi-Command”.
Summary of Contents for HARS Series
Page 1: ...Instruction Manual F RA 001 3R3 MODEL HARS series B N 277 9 002 277 9 002 Rev 0 1 ...
Page 10: ......
Page 19: ...2 Before Using This Product 9 HARS series 2 6 3 Dimensions a Models 500W 1kW 60kV or lower ...
Page 20: ...2 Before Using This Product 10 HARS series b Models 500W 1kW 100kV 120kV ...
Page 21: ...2 Before Using This Product 11 HARS series c Models 2kW 3kW 60kV or lower ...
Page 22: ...2 Before Using This Product 12 HARS series d Models 2kW 3kW 100kV 120kV ...
Page 92: ......