3 Connecting
15
HARS series
3-2 Connecting the GND
For safety, be sure to ground the GND terminal of the unit at a single point.
3-2-1 No grounding for the load frame (Models at rated current 1.3A or higher)
3-2-2 No grounding for the load frame (Models at rated current 1.3A or lower)
Rack-mount case
Ground
Output cable
GND terminal
Load
frame
Black cable (M6)
▪ 5SQ or more
▪ Within about 2 meters
▪ Do not use bare wire.
Class D (Class three) grounding at
single point
Rack-mount case
Ground
Output cable
GND terminal
Load
frame
▪ 5SQ or more
▪ Within about 2 meters
▪ Do not use bare wire.
Class D (Class three) grounding at
single point
Load
Load