RU-200,RU-500

FC-200 and FC-500

Refrigerated units

USER’S MANUAL

Issue 11.

Date of  issue:

3rd May 2005

Techne is a trademark

© Techne, 2003

Summary of Contents for Techne FC-200

Page 1: ...RU 200 RU 500 FC 200 and FC 500 Refrigerated units USER S MANUAL Issue 11 Date of issue 3rd May 2005 Techne is a trademark Techne 2003...

Page 2: ......

Page 3: ...INTRODUCTION 8 Dip Coolers 8 Flow Coolers 9 Operator Safety 9 Specification 10 Siting Considerations 11 Setting up the unit 11 Carrying a Dip Cooler 14 Cleaning your unit 14 Packing 14 OPERATION 15 Pr...

Page 4: ...P A G E 2 D I P A N D F L O W C O O L E R S U S E R S M A N U A L...

Page 5: ...P A G E 3 D I P A N D F L O W C O O L E R S U S E R S M A N U A L NOTES...

Page 6: ...nings I mains switch On O mains switch Off After use When you have finished heating samples remember that partsoftheunit thetubes blocksandassociatedaccessories may be very hot Take the precautions li...

Page 7: ...an de travail ou surface plane ou le cas ch ant dans une hotte d aspiration en s assurant que les trous d a ration situ s sous l appareil ne sont pas obstru s 5 Les symboles situ s sur ou c t de l int...

Page 8: ...bene Arbeitsfl che bzw fallserforderlich untereinenLaborabzug BeachtenSie da dieEntl ftungsrippenanderGer teunterseiteimmer frei zug nglich sind 5 Die Symbole auf oder neben dem EIN AUS Schalter an de...

Page 9: ...tradas de aire en la parte inferior no queden obstruidas 5 Loss mbolosqueseencuentranenocercadelinterruptor de alimentaci n tienen los siguientes significados I Interruptor principal encendido O Inter...

Page 10: ...this could be a Techne Thermoregulator TE 10 series or TU 20 series or it could be a Techne Circulator the C 85A or C 85D Each unit fridge or controller has its particular application which should be...

Page 11: ...d on the rear of the unit These pipes are 16mm diameter The flow coolers need a minimum flow rate of 10 litres min at 20o C to prevent freezing in the heat exchanger OUTLET INLET Operator Safety All u...

Page 12: ...COOLERS RU 200 RU 500 Minimum temperature 20 o C 35 o C Cooling capacity at 20o C 145 W 240 W 0o C 145 W 240 W 10 o C 110 W 230 W Nominal dimensions and weights Overall Length 420 mm 430 mm Width 235...

Page 13: ...nce MINIMUM MINIMUM 150 mm 6 150 mm 6 Setting up the unit 1 Check the voltage marked on the unit s rating plate against that of the mains supply Connect the mains cable to a suitable plug as in the fo...

Page 14: ...ch will cover the full range of working temperatures The units are designed to work up to 40o C with occasional peaks to 100o C see Precautionary Measures in the Operation section However with the coi...

Page 15: ...hose may be used if the hose is soft such as silicone rubber Ensure the flow is in through the INLET pipe and out through the OUTLET as cooling power will be lost if the flow is reversed OUTLET INLET...

Page 16: ...r to give the support stability The coil is then clipped into the support for ease of carrying RU 500 RU 200 Cleaning your unit Before cleaning your unit disconnect from the power supply and allow to...

Page 17: ...with liquids above 40o C 4 Although the practice is not recommended fridge units may occasionally be subjected to temperatures up to 100o C This is provided that they are only being used thus intermi...

Page 18: ...the liquid is not being circulated over the coil or through the heat exchanger properly First check that the pump unit is not blocked either by foreign matter or by balls from a ball blanket or in any...

Page 19: ...gas charge for each of the units Unit Refrigerant Charge RU 200 R134a 95 gm RU 500 R404a 250 gm FC 200 R134a 95 gm FC 500 R404a 240 gm There have been several improvements in design which have varied...

Page 20: ...details of the fault observed and remember to return the unit in its original packing Techne accept no responsibility for damage to units which are not properly packed for shipping if in doubt contac...