Masimo Root RDS-7 Operator'S Manual Download Page 1

 

 

 

 

 

Root

 

Operator's Manual 

 

 

Summary of Contents for Root RDS-7

Page 1: ...Root Operator s Manual ...

Page 2: ......

Page 3: ...his device to sale by or on the order of a physician For professional use See instructions for use for full prescribing information including indications contraindications warnings precautions and adverse events Wireless Radio IC 7362A RDS7 IC Model RDS 7 FCC ID VKF RDS7 Model RDS 7 Masimo Corporation 40 Parker Irvine CA 92618 USA Tel 949 297 7000 Fax 949 297 7001 www masimo com EU authorized repr...

Page 4: ...o com 2 Masimo The use of the trademarks PATIENT SAFETYNET and PSN is under license from University HealthSystem Consortium All other trademarks and registered trademarks are property of their respective owners ...

Page 5: ...ings and Cautions 13 Electrical Information Warnings and Cautions 15 Chapter 1 Description 17 Features 17 Chapter 2 Setting Up 21 Unpacking and Inspection 21 Guidelines for Setting Up 21 Power On 22 Initial Battery Charging 23 Radical 7 Connection 24 Nurse Call Connection Port 24 Chapter 3 Operation 27 About the Main Screen 28 About the Status Bar 29 Using the Touchscreen Interface 31 Menu Navigat...

Page 6: ...on Ports 70 Iris Status Screen 71 Chapter 6 Messages 73 Chapter 7 Troubleshooting 75 Troubleshooting Radical 7 and MOC 9 Modules 75 Troubleshooting Root 75 Chapter 8 Specifications 79 Electrical 79 Environmental 79 Display 80 Alarms 80 Nurse Call Specifications 81 Connectors 81 Communication Specifications 81 Applicable Standards and Regulations 82 Symbols 84 Chapter 9 Service and Maintenance 87 C...

Page 7: ...Replacing the Fuses 88 Power On Self Test 88 Nurse Call Setting Connections 88 Battery Test 90 Repair Policy 90 Return Procedure 91 Contacting Masimo 91 Sales End User License Agreement 92 Warranty 92 Exclusions 93 Restrictions 93 Index 95 ...

Page 8: ......

Page 9: ...n a serious outcome for example injury serious adverse effect and death to the patient or user The following is an example of a warning WARNING This is an example of a warning statement A caution is given when any special care is to be exercised by the patient or user to avoid injury to the patient damage to Root or damage to other property The following is an example of a caution CAUTION This is ...

Page 10: ......

Page 11: ...pansion through Masimo Open Connect MOC 9 Designed for third party measurement expansion to allow other companies to add to the platform measurements Built in connectivity gateway through Iris for standalone devices such as IV pumps ventilators beds and other patient monitors Intended Use Root is a docking station for the Radical 7 handheld monitor It is also a platform device intended for use wit...

Page 12: ......

Page 13: ...RNING Do not place Root or accessories in any position that might cause it to fall on the patient WARNING Do not use Root during magnetic resonance imaging MRI WARNING Explosion hazard Do not use the Root in the presence of flammable anesthetics or other flammable substance in combination with air oxygen enriched environments or nitrous oxide WARNING Replace only with fuses of same type current ra...

Page 14: ... based solutions acetone solutions or other harsh solvents to clean Root These substances affect the device s materials and device failure can result CAUTION Ensure that the speaker is not covered CAUTION Do not place Root against a surface that may cause the alarm to be muffled CAUTION Should the Radical 7 and Root become unable to communicate parameters and measurements will not show on the Root...

Page 15: ...accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense In order to maintain compliance with FCC regulations shielded cables must be used with this equipment Operation with non approved...

Page 16: ... Information Warnings and Cautions www masimo com 14 Masimo Note This Class A digital apparatus complies with Canadian ICES 003 Note Use Root in accordance with Environmental Specifications section in this manual ...

Page 17: ...ry out periodic tests to verify that leakage currents of patient applied circuits and the system are within acceptable limits as specified by the applicable safety standards The summation of leakage currents must be checked and in compliance with IEC 60601 1 and UL60601 1 The system leakage current must be checked when connecting external equipment to the system When an event such as a component d...

Page 18: ...lectrical outlet controlled by a wall switch or dimmer CAUTION Failure to plug in the AC power cable promptly after a low battery message may result in the device shutting down Note Disconnect the instrument using the AC power cord connector Note It is recommended that Root is attached to an AC power source when it is not in use to ensure that the battery remains fully charged Note For medical tec...

Page 19: ...ivity gateway for standalone devices Features Front View Ref Feature Description 1 Docking Station Provides a docking station for the Radical 7 While docked the Radical 7 can communicate monitored parameters and measurements 2 Root Display and Touchscreen Provides an interface for user interactions 3 Home Button Provides access to the Main Screen 4 Root Charging Indicator Shows an indication of th...

Page 20: ...for the Radical 7 in the Docking Station Only the touchscreen version of the Radical 7 is able to communicate monitored parameters and measurements All other versions can only charge in the docking station but not communicate with Root Back View Feature Description 1 Handle Allows the user to transport Root 2 Speaker Provides audible notification 3 Nurse Call Connector Provides a connection to a N...

Page 21: ...ble 5 USB Ports 2 Provide USB 2 0 connectivity 6 Power Entry Module Contains the input connector for a hospital grade AC power cord and the fuse holder 7 Equipotential Ground Connector Provides electrical grounding for Root 8 Iris Connectivity Ports 4 Provide connection for standalone devices Side Views Right side Left side Feature Description ...

Page 22: ...Feature Description 1 System Status Lights Provides an indication of alarms and system messages See System Status Lights on page 63 2 Power Button Places Root in Power On Standby and Power Off modes 3 MOC 9 Ports 3 Provide connectivity for MOC 9 modules ...

Page 23: ...epartment See Return Procedure on page 91 Guidelines for Setting Up Root has a built in bracket interface that allows it to be mounted on a pole or roll stand When setting up Root follow these guidelines Place on a stable hard flat and dry surface near the patient Maintain a minimum of three 3 centimeters one 1 inch of free space around Root Ensure that the back panel Speaker is not covered to avo...

Page 24: ... Standby and Power Off To Power On press the Power Button for two 2 seconds until a single audible tone sounds Once Root turns on if no Radical 7 or MOC 9 module is connected the Root display shows the following message Please Connect a Device The user is now able to connect Radical 7 and MOC 9 modules ...

Page 25: ...C power source 3 Verify that Root s battery is charging by ensuring that the AC Power Indicator 1 is green and the Battery icon on the Status Bar 2 is solid green or has the charging symbol See AC Power Indicator on page 63 and About the Status Bar on page 29 1 2 4 The Root Charging Indicator remains orange while the battery is charging and will illuminate green when Root is fully charged See Batt...

Page 26: ...ss the power button on the Radical 7 to power it on When properly connected the Radical 7 Charging Indicator light will illuminate An illuminated Radical 7 Battery icon will also appear in the Status Bar Root display will show active measurements and parameters For Radical 7 charging conditions see Radical 7 Charging Indicator on page 64 Nurse Call Connection Port Use a Nurse Call connection cable...

Page 27: ...male connector of the cable 2 Insert the Nurse Call connection cable securely into the compatible port on Root 3 Depending on the connection type of the Nurse Call System it may be necessary to orient the other end of the Nurse Call connection cable to fit correctly into the system connection 4 For more information see Device Output on page 51 ...

Page 28: ......

Page 29: ...on The information in this chapter assumes that Root is set up and ready for use This chapter provides necessary information for proper operation of the device Do not operate Root without completely reading and understanding these instructions ...

Page 30: ...ography measurements and SedLine module bottom showing brain function measurements Ref Feature Description 1 Status Bar Displays system status as well as icons that provide shortcuts to menu items or actions See About the Status Bar on page 29 2 Windows Provides a dynamic user configurable display area for all the data from Radical 7 or a MOC 9 module 3 Main Menu icon Provides access to the config...

Page 31: ...ee Audio Pause on page 62 3 Profiles Provides access to the Profiles screen The example shown illustrates that Profiles is currently set to Adult for an adult patient See Profiles on page 54 4 Bluetooth Provides access to the Bluetooth screen If this icon is visible then Bluetooth connectivity has been enabled See Bluetooth on page 47 5 Wi Fi Provides access to the Wi Fi screen If this icon is vis...

Page 32: ...e Battery Radical screen The example shows that the battery is currently charging See Radical 7 Charging Indicator on page 64 9 Root Battery Displays charging status for Root and provides access to the Battery Root screen The example shows that the battery is currently charging See Root Charging Indicator on page 64 10 Sounds Provides access to the Sounds screen to adjust alarm and pulse tone volu...

Page 33: ...ouch move left right up or down and release Moves an object across the display Swipe Touch move left right up or down and release quickly Pinch Touch move and release via two touch points Moving touch points apart zooms in and moving them together zooms out Drag and Drop See Customizing Windows on page 41 Touch hold drag an object to desired position and drop it by releasing Below is a list of all...

Page 34: ...pen scrolls through spinner tiles Press unfocused tile When open scrolls tile into center focused position Press anywhere outside spinner When open collapses spinner Slider Slide knob Moves knob Press anywhere along slider path Quickly moves knob to Tap position Slider Spinner Slide knob Moves knob Press anywhere along slider path Quickly moves knob to Tap position Press center focused tile When c...

Page 35: ...measurement When no parameter or measurement alarm is present opens parameter or measurement menu When parameter or measurement alarm is present silences parameter or measurement alarm Press and hold Enables parameter and measurement drag and drop Well Press parameter or measurement When no parameter or measurement alarm is present opens parameter or measurement menu When parameter or measurement ...

Page 36: ...label Resets time range to default Alarm Silence icon Press Silences all alarms Audio Pause icon Press Enables Audio Pause Other Status Bar icons Press Opens relevant menu Back Arrow Press Exits menu abandons any changes Menu Navigation When navigating through menus and configuring settings all changes must be confirmed by selecting OK To cancel the changes select Cancel Any screen requiring selec...

Page 37: ...ers or measurements can be expanded or minimized within a Window to customize view Radical 7 Windows are shown in the examples below Windows provide waveforms along with either a Trend View or an Analog View Trend View displays each parameter or measurement alongside a graph of its values over time Analog View displays values in relation to alarm ranges Details about the displayed information of p...

Page 38: ...play includes its Value Range Numeric Value Alarm Limits and Parameter label See Using Trend View on page 37 4b Analog Gauge Available only in Analog View Parameters and measurements are shown as Analog Gauges in Analog View A parameter s Analog Gauge includes its Alarm Limits Numeric Value Parameter Label as well as Alarming Caution and Normal Ranges See Using Analog View on page 38 5 Time Frame ...

Page 39: ...end Display in Trend View Feature Description 1 Value Range Indicates current viewing of the parameter or measurement Press to access the Trend menu where the minimum and maximum of the range can be modified 2 Trend Line Displays parameter and measurement over a period of time Zoom in and out of a Trend Line by pinching out and in 3 Numeric Value Indicates current reading of the parameter or measu...

Page 40: ...rator s manuals for Radical 7 and the appropriate MOC 9 module s If a Window supports Analog View then Analog View icon displays in the upper right corner Press the icon to view parameter and measurement data in Analog View The following diagrams and tables describe key features of a parameter s Gauge in Analog View When alarm limits for a specific parameter or measurement are set the correspondin...

Page 41: ...Root Chapter 3 Operation www masimo com 39 Masimo Ref Feature Description 4 Parameter or Measurement Label Indicates the name of the parameter or measurement ...

Page 42: ...alarm will be triggered Some ranges display as quarter circles others display as half circles A quarter circle displays when the value has a physiologic normal level at one end of the range A half circle displays when the value has a physiologic normal level in the middle of the display range In the example below the SpO2 gauge is shown as a quarter circle where values lower than 88 will trigger a...

Page 43: ...View When a parameter is minimized it is only displayed the Well with its Numeric Value and Parameter Label When a parameter is expanded it will be shown as either a Trend Display or Gauge To expand a parameter or measurement Order Instruction Step 1 Press and hold the Numeric Value until it dims Step 2 Drag the Numeric Value over any Trend Display Step 3 Release the Numeric Value ...

Page 44: ...ration www masimo com 42 Masimo Minimizing a parameter or measurement Order Instruction Step 1 Press and hold the Numeric Value until it shrinks Step 2 Drag the Numeric Value to the Well Step 3 Release the Numeric Value ...

Page 45: ...tom right corner of the touchscreen press the Main Menu icon The Main Menu options are Device Settings See Device Settings on page 44 About See About on page 53 Trend Settings See Trend Settings on page 53 Profiles See Profiles on page 54 Iris See Iris on page 54 Layout See Layout on page 55 Sounds See Sounds on page 57 ...

Page 46: ...ngs options are Localization See Localization on page 45 Wi Fi See Wi Fi on page 46 Ethernet See Ethernet on page 46 Bluetooth See Bluetooth on page 47 Battery Root See Battery Root on page 47 Battery Radical See Battery Radical on page 48 Brightness See Brightness on page 49 Access Control See Access Control on page 50 Device Output See Device Output on page 51 ...

Page 47: ...ee About the Status Bar on page 29 Option Description Factory Default Setting Configurable Settings Language Selects the language display for Root English Choose from available languages Date Format Sets the display format for current date mm dd yy mm dd yy or dd mm yy Time Sets the display format for current time 12 hour 12 hour or 24 hour Line Frequency Sets to match regional power line frequenc...

Page 48: ...ly settings about the Wi Fi connection that cannot be configured by the user Your Masimo sales representative can provide necessary information regarding an initial Wi Fi connection Ethernet Use the Ethernet screen to enable or disable Ethernet connectivity When Ethernet connectivity is enabled the Ethernet icon will appear in the Status Bar See About the Status Bar on page 29 Option Description F...

Page 49: ...g Configurable Settings Bluetooth Enables or disables Bluetooth connectivity Off On or Off Battery Root Use the Battery Root screen to view the specific percentage of charge on the battery The user can also access Root s Battery screen by pressing the Battery icon on the Status Bar See About the Status Bar on page 29 Option Description State of Charge Provides a read only display of battery level ...

Page 50: ...e Radical 7 s battery The user can also access the Battery screen by pressing the Radical 7 s Battery icon on the Status Bar See About the Status Bar on page 29 Option Description State of Charge Provides a read only display of battery level remaining Battery Diagnostics Allows trained personnel to access battery diagnostic information ...

Page 51: ...htness of the Root display Option Description Factory Default Setting Configurable Settings Auto Brightness Allows automatic adjustment of Root s display brightness based on ambient light Off On or Off Brightness Adjust the brightness level of the Root display by sliding the button 4 is brightest 4 1 2 3 4 ...

Page 52: ...ed To undo an entry press Backspace 3 Press Enter to access the password protected screen Note The password will have to be entered every time this screen is accessed Option Description Factory Default Setting Configurable Settings All Mute Enabled Enable parameter Alarm Silence menu option See Sounds on page 57 Off On or Off Lock Alarm Volume Sets the lowest alarm volume level Off 3 4 or Off Save...

Page 53: ...t Device Output configures the Nurse Call feature on Root A Nurse Call can be triggered based on alarm low Signal IQ events or both In addition Nurse Call Polarity can be inverted to accommodate local Nurse Call station requirements Option Description Factory Default Setting Configurable Settings Nurse Call Polarity Controls the mechanism of action for triggering to occur Should be changed to acco...

Page 54: ...r 3 Operation www masimo com 52 Masimo Note The Nurse Call feature is disabled when Audio Pause is enabled and Nurse Call Trigger is set to Alarms For more information about Audio Pause see Audio Pause on page 62 ...

Page 55: ...rsion number MCU 2 Displays software version number Information about Radical 7 and MOC 9 modules will display in a separate list These fields are read only and cannot be configured by the user Trend Settings Use the Trend Settings screen to configure trend viewing on the Main Screen and trend data storage on Root Option Description Factory Default Setting Configurable Settings Default Duration Du...

Page 56: ...tion controls the management of patient configuration settings on Root The settings of the three default profiles Adult Pediatric and Neonatal configure parameter alarms averaging time and sensitivity modes Custom profiles can also be created to accommodate usage in any hospital care area See the Masimo Instrument Configuration Tool Directions for Use for information on adding Profiles For more in...

Page 57: ...play the parameters and measurements by changing the Layout Style Note This feature only applies to Trend View See Using Trend View on page 37 Option Description Factory Default Setting Configurable Settings Layout Style Controls the sizing of Trend Displays Fixed Fixed or Dynamic The following diagram and tables explain the differences between Fixed and Dynamic modes for a Trend View ...

Page 58: ...indows on page 41 2 Size of each Trend Display is fixed Dynamic Ref Description 1 Size of all Trend Displays decreases or increases to accommodate parameter s expanded or minimized All Trend Displays are always evenly sized For more information about expanding and minimizing parameters see Customizing Windows on page 41 2 Size of each Trend Display is automatically adjusted When the number of Tren...

Page 59: ...ilable Layout When only Radical 7 or a single MOC 9 module is connected to Root the Available Layout will be 100 When Radical 7 and or multiple MOC 9 modules are connected the user will have the option to select from several pre configured layouts Sounds ...

Page 60: ...ns silenced when Audio Pause is enabled See Audio Pause on page 62 2 minutes 1 2 3 minutes Permanent Permanent with Reminder If Permanent is selected there will be no audible alarms but visual alarms will still display If Permanent with Reminder is selected a tone will sound every three 3 minutes as a reminder that Permanent is active Requires user to have All Mute Enabled Permanent Silence Durati...

Page 61: ...te that the borders of both the PR Trend Display as well as the Window are illuminated red and the explanation of the alarm is shown at the top of the Window High PR 140 Window Level The example shown here is a High Impedance alarm in the lower Window Note that the border of the Window illuminates red and the explanation of the alarm is shown at the top of the Window High Impedance Detected ...

Page 62: ...adical 7 and MOC 9 modules Alarm Silence The Alarm Silence icon is an indicator as well as a functional button It always indicates the presence of alarms and it can be used to temporarily suspend audible alarms for a pre configured amount of time known as Silence Duration Silence Duration configurations vary across different parameters and measurements For more information about Silence Duration r...

Page 63: ...ption Visual Alarms There are currently no active alarms but at least one alarm has been and is still silenced No There is currently at least one active alarm that has not been silenced Yes There is currently at least one active alarm but all active alarms are silenced Yes ...

Page 64: ...udio Pause is 120 seconds In the example below Audio Pause is activated and there are 117 seconds left until Audio Pause is inactivated again To configure Audio Pause see Sounds on page 57 Audio Pause active 117 seconds until Audio Pause is inactive Note When Audio Pause is activated powering off and then powering on Root will return Audio Pause to its default inactive state Trend Download Root ca...

Page 65: ...tus Alarm Priority Indication None None Monitoring has not begun Green None There is currently no active alarms Yellow Low There is an active alarm of low priority Red High There is an active alarm of high priority The alarm priority is determined by the Radical 7 and MOC 9 module s that are connected to Root The following are system level alarm messages that accompany System Status Lights when Ra...

Page 66: ...ource Off Root is not connected to an AC power source Root Charging Indicator Whenever Root is connected to an AC power source if not fully charged its battery will charge Light Status Battery Indication Green Battery is fully charged Orange Battery is charging Red Battery charging error Off Battery is not being charged Root is not connected to AC power source Radical 7 Charging Indicator ...

Page 67: ...ully charged Orange Battery is charging Red Battery is unable to charge Off Battery is fully charged not being charged Standby and Power Off To put Root in the Standby Mode or Power Off follow these steps State Description Standby Mode Press and hold the Power Button for two 2 seconds until one 1 audible tone sounds Standby Mode conserves power while enabling a quicker startup sequence Power Off P...

Page 68: ......

Page 69: ...plays all the data from that module The example below shows the SedLine and Capnography Windows which display data from the SedLine brain function monitoring and Phasein capnography MOC 9 modules that are connected to Root MOC 9 Ports Use a MOC 9 cable to connect other MOC 9 modules to Root To use an MOC 9 Port 1 Identify the MOC 9 end of the cable 2 Orient the cable to fit correctly into an MOC 9...

Page 70: ......

Page 71: ...ronic health records Below is an example of one way Root can be used in a network setting using Patient SafetyNet Root receives and may display information from Radical 7 MOC 9 modules as well as standalone devices Ref Description 1 4 Standalone devices connected via Iris e g monitor pump ventilator 5 Root 6 Radical 7 7 MOC 9 modules 8 Patient SafetyNet or Masimo Connectivity Gateway 9 Electronic ...

Page 72: ... standalone devices to Root Port 1 Port 2 Port 3 Port 4 To connect a standalone device via an Iris Connectivity Port 1 Connect the RS 232 end of the Iris Adapter to the standalone device 2 Connect the RJ 45 end of the Iris Adapter to any of the four 4 compatible Iris Connectivity Ports on Root using a RJ 45 cable ...

Page 73: ...afetyNet or Connectivity Gateway Yellow Standalone device is successfully connected to Root however Root is not successfully connected to a Patient SafetyNet or Connectivity Gateway Gray No standalone device is connected to the Iris Connectivity Port Additionally the Ports are also mapped to the Iris icon on the Status Bar When a standalone device is connected to Root via one of the Ports the corr...

Page 74: ...mo Port 3 Port 4 Status and connection type are read only and not configurable by the user For more information on Iris connectivity refer to the instructions for use or operator s manual for the appropriate version of Masimo Patient SafetyNet ...

Page 75: ...ery using AC power MOC 9 module disconnected e g SedLine Disconnected A MOC 9 module is disconnected from Root Reconnect module or acknowledge message by pressing the Alarm Silence icon Radical 7 Disconnected Radical 7 is disconnected from Root Reconnect Radical 7 or acknowledge message by pressing the Alarm Silence icon For additional messages refer to the instructions for use or operator s manua...

Page 76: ......

Page 77: ...attery One of the fuses is not operating properly Replace the fuse See Replacing the Fuses on page 88 Root turns on but Main Screen is dim or blank The brightness setting is not correct Adjust the brightness setting See Brightness on page 49 If the condition persists Root requires service Contact Masimo Technical Support See Return Procedure on page 91 Touch functionality is not responsive Interna...

Page 78: ...tinuous speaker tone Internal failure To silence an alarm press the Power Button for eight 8 seconds If alarm continues to sound Power Off Root Root requires service See Return Procedure on page 91 Power Button does not respond when pressed Power Button may need to be pressed for a longer time To Power On when turned off or in Standby Mode press Power Button for two 2 seconds To Power Off when tur...

Page 79: ...ssible Cause Correction Root Charging Indicator illuminates red Internal failure Root requires service See Return Procedure on page 91 Nurse Call does not communicate Connection error Unplug and replug Nurse Call connector See Nurse Call Setting Connections on page 88 ...

Page 80: ......

Page 81: ...er requirements 100 240 VAC 47 63 Hz 65 W Max Fuses 2 2 Amp Fast Acting Metric 5x20mm 250V Battery Type 10 8V Lithium Ion Nominal Capacity 4 hours Maximum Charging Time 4 hours This represents approximate run time at the lowest brightness using a fully charged battery Environmental Root Operating Temperature 32 F to 122 F 0 C to 50 C Transport Storage Temperature 40 F to 158 F 40 C to 70 C Operati...

Page 82: ...8 000 ft 304 m to 5 486 m Display Characteristic Description Type Backlit Active Matrix TFT LCD Resolution 1280 x 800 pixels Size 10 1 in 25 65 cm Diagonal Color 24 bit RGB Touchscreen Characteristic Description Type Multi Touch P Cap Alarms Visual Alarm Priority System Status Light Color High Red Low Yellow ...

Page 83: ...tion Max Voltage 36 VDC or 24 VAC peak Connectors Connector Type Number of Ports Ethernet 10 100 MBps 1 Nurse Call 1 4 inch round female 1 MOC 9 Masimo Connector 3 USB USB 2 0 2 Iris RS 232 RJ 45 4 Communication Specifications Communication Type Description WLAN Radio Tri Mode IEEE 802 11 a b g Security and Authentication Encryption 64 128 bit WEP Dynamic WEP WPA TKIP WPA2 AES ...

Page 84: ...N 60601 1 4 1996 A1 1999 EMC Compliance EN 60601 1 2 Class A device ISO 9919 Clause 36 20 V M radiated immunity Protection from Ingress of Liquids IPX1 per IEC 60529 Electrical and Mechanical Safety UL 60601 1 2003 CAN CSA 22 2 No 601 1 2005 IEC 60601 1 UL 60601 1 1988 A1 1991 A2 1995 EN 60601 1 1990 AI 1993 A2 1995 ISO 9919 2005 Type of Protection Class I on AC power Internally powered on battery...

Page 85: ...ns www masimo com 83 Masimo Compatibility Standards USA FCC ID VKF RDS7 Model RDS 7 FCC parts 15 207 15 209 15 247 and 15 407 Canada IC 7362A RDS7 IC Model RDS 7 RSS 210 Europe R TTE Directive EN 3000 328 EN 301 893 EN 301 489 17 ...

Page 86: ...Caution consult accompanying documents WEEE Compliant Manufacturer Storage temperature limitation Storage altitude limitation Date of Manufacture Storage Humidity Limitation Non ionizing electromagnetic radiation Keep dry IPX1 Protection against liquid drops falling vertically Fragile handle with care Federal USA law restricts this device to sale by or on the order of a physician Fuse replacement ...

Page 87: ...Iris Connection FCC ID IC IC Model Identifies unit has been registered as a radio device USB port Mark of conformity to European Medical Instrument Directive 93 42 EEC AC Current Authorized representative in the European community Nurse Call interface Wireless features can be used in member states with the restriction of indoor use in France ...

Page 88: ......

Page 89: ...ranty Cleaning Root is a non sterile and reusable device To surface clean wipe down the outer surface of Root using any of the following solvents Cidex Plus 3 4 glutaraldehyde 10 bleach solution 70 isopropyl alcohol solution Do not allow liquids to enter the interior of Root Using the recommended cleaning solutions on the touchscreen will not affect the performance of Root ...

Page 90: ... Place a new fuse in the fuse holder 6 If replacing both fuses repeat steps 4 and 5 for the second fuse 7 Slide the fuse holder back into the Power Entry Module and press firmly to make sure it is secure Root is ready to be reconnected to AC power If the fuses fail shortly after replacement Root requires service See Repair Policy on page 90 Power On Self Test To conduct a Power On Self Test 1 Conn...

Page 91: ...tal s Nurse Call system Cable Description Nurse Call Event Menu Setting 2 contacts normally opened Nurse Call Polarity Normal 2 contacts normally closed Nurse Call Polarity Inverse 1 and 2 contacts normally opened 2 and 3 contacts normally closed Nurse Call Polarity Normal 1 and 2 contacts normally closed 2 and 3 contacts normally opened Nurse Call Polarity Inverse 1 and 2 contacts normally closed...

Page 92: ...On Root and verify that the Root Battery Indicator icon on the Status Bar shows a full charge Repair Policy Masimo or an Authorized Service Provider must perform warranty repair and service Have the malfunctioning device repaired Clean a contaminated and or dirty device before returning following the cleaning procedure described in Cleaning on page 87 Make sure the device is fully dry before packi...

Page 93: ...nical Services Warranty or extended warranty information and a copy of the invoice or other applicable documentation must be included Purchase order number to cover repair if the device is not under warranty or for tracking purposes if it is Ship to and bill to information Contact information name telephone fax number and country for any questions about the repair A certificate stating the product...

Page 94: ...ages with your sales receipt to Masimo for a full refund Warranty Masimo warrants to the initial Purchaser for a period of one 1 year from the date of purchase that i each new Product and the Software media as delivered are free from defects in workmanship or materials and ii the Product and Software will perform substantially as labeled in the directions for use Masimo s sole obligation under thi...

Page 95: ...ct and Software This warranty is expressly in lieu of any oral or implied warranties including without limitation any implied warranty of merchantability or fitness for a particular purpose Masimo shall not be liable for any incidental special or consequential loss damage or expense directly or indirectly arising from the use or loss of use of any Products or Software In no event shall Masimo s li...

Page 96: ...title to the Product with which this Software is included Any attempt to assign any rights duties or obligations arising hereunder other than as set forth in this paragraph shall be void 4 U S Government Rights If Purchaser is acquiring Software including the related documentation on behalf of any part of the United State Government the following provisions apply the Software and documentation are...

Page 97: ...9 Bluetooth 27 41 44 Brightness 41 46 69 C Chapter 1 Description 17 Chapter 2 Setting Up 21 Chapter 3 Operation 25 Chapter 4 MOC 9 61 Chapter 5 Iris 63 Chapter 6 Messages 67 Chapter 7 Troubleshooting 69 Chapter 8 Specifications 71 Chapter 9 Service and Maintenance 77 Cleaning 14 77 79 Communication Specifications 73 Compliance Information Warnings and Cautions 13 Connectors 73 Contacting Masimo 80...

Page 98: ...l Specifications 72 P Power On 22 Power On Self Test 77 Product Description 9 Product Description and Intended Use 9 Profiles 27 40 50 R Radical 7 Charging Indicator 24 28 59 Radical 7 Connection 24 Repair Policy 77 79 Replacing the Fuses 69 77 Restrictions 81 Return Procedure 21 69 70 79 Root Charging Indicator 28 59 S Safety Information Warnings and Cautions 11 23 Sales End User License Agreemen...

Page 99: ...Root Index www masimo com 97 Masimo U Understanding Windows 33 Unpacking and Inspection 21 Using Analog View 34 36 Using the Touchscreen Interface 29 Using Trend View 34 51 W Warranty 80 Wi Fi 27 41 43 ...

Page 100: ......

Page 101: ......