Additional Settings
Use the
Additional Settings
screen to configure the following:
Options*
Description
Factory Default Settings User Configurable Settings
Apnea Timeout**
Displays the set no-breath timeout 20 seconds
NA
Set O
2
Range**
Displays the set O
2
range
0-30 vol%
NA
Set N
2
O Range**
Displays the set N
2
O Range
0-30 vol%
NA
CO
2
Unit of Measure Sets the CO
2
display unit
mmHg
mmHg, kPa***, or vol%
* Capnography setting changes made on the Masimo device are not transferred to the Radius PCG device. Changes to
Capnography settings for the Radius PCG device are made on the device itself. See
Chapter 4: Operation
on page 27
for information about settings for the Radius PCG device.
** These fields are read-only and cannot be configured by the user.
*** Selecting kPa will cause all stored Trend data to be erased.