Chapter 6: Troubleshooting
Troubleshooting Radius PCG
Error
Possible Causes
Recommended Solutions
No Adapter alarm is displayed
Indicates that an EMMA Airway
Adapter needs to be installed.
Connect an EMMA Airway Adapter. See
Attaching Airway Adapter
on page 23.
Clogged Adapter alarm is
displayed
Indicates that the EMMA Airway
Adapter needs to be replaced with
a new one.
Replace the EMMA Airway Adapter with a new
one.
Zero point adjustment alarm is
displayed
A zero point adjustment is
required.
Perform a Zeroing procedure. See
Zeroing
Procedure
on page 60.
The unit does not complete the
turn on sequence
Low battery
Replace the batteries.
The unit does not turn on
•
No battery
•
Low battery
Replace the batteries.
The measured values of EtCO
2
are out of specified accuracy
Incorrect zero reference
Perform a Zeroing procedure and verify the
measurement with reference gas. See
Chapter
8: Service and Maintenance
on page 57.