background image

Doc.Ref. No. m83B/QG/101                      Issue No.:00                                                  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

                       Page 

1

 of 

2

 

 

 

 

 

 

85XX+ Scanner/DAQ

 

 

Input 

Type  

Range 

Input 

Type 

Range 

-200 to  1000

CU53 

-210.0 to 

210.0°C 

-200 to  1200

NI120 

-70.0 to 
210.0°C 

-200 to  1370

C  -10 to 20 

mV 

-1999 

to 9999  

-200 to  400

0 to 100 

mV 

450 to 1800

4-20mA  

0 to 1750

0 to 20 

mA 

0 to 1750

0 – 5 V 

-200 to  1300°C  

1 – 5 V 

RTD(PT100)  -199.9 to 850.0

0 – 10V 

Table 1.1 

 

SPECIFICATIONS 

 

NO. OF CHANNEL 

8 or 16 or 24 

ACCURACY 

TC/RTD/LINEAR: 

 
+ 0.1% of instrument 

range + 1 digit

 

RESOLUTION 

TC(E,J,K,T)/RTD/CU53/NI120: 
TC(B,R,S,N): 
LINEAR: 

ADC: 17 bits 

0.1

1

1 Count 

APPLICABLE STANDARD 

DIN (ITS-90) for 
Thermocouple and RTD 

INPUT TYPE 

Refer table 1.1 

SAMPLING PERIOD PER 
INPUT 

50 ms for TC and Linear 
Input and 100 ms for 
RTD 

BURNOUT CURRENT 

0.4 µA 

MEASUREMENT CURRENT 

250 µA 

INPUT IMPEDANCE 

>1 MΩ for RTD/Voltage 
inputs, 250 Ω for 
current Input 

NMRR 

> 40 dB (50/60 Hz) 

CMRR 

>120 dB (50/60 Hz) 

ALLOWABLE WIRING 
RESISTANCE FOR RTD 

Maximum 15 ohms/wire 
(Conductor resistance 
between three wires 
should be equal). 

Digital Input Specification (Optional)*

 

* With Digital Input, CE marking is not applicable/valid 
 

Digital Output- Relay

 

NUMBER OF OUTPUTS 

PURPOSE  

Alarm or trip or control or 
watchdog output  

OUTPUT SIGNAL 

Two terminals (C and NO)  

RELAY CONTACT 
RATING 

250 VAC / 30 VDC @ 2A 

NUMBER OF RELAY 
OPERATION 

1 X 10^5 @ rated current 

 

Digital Output- Open Collector (Optional)

 

NUMBER OF OUTPUTS 

24 

PURPOSE  

Alarm or trip or control or 
watchdog output 

OUTPUT TYPE 

transistor open collector 
output selection) 

CONTACT RATING 

30 V DC,100 mA 

 

Analog Output- Analog Output (Optional)*

 

NUMBER OF OUTPUTS 

OUTPUT SIGNAL 

0-20 mA, 4-20 mA or 0-5 V, 
1-5 V, 0-10 V DC 

LOAD RESISTANCE 

For current o/p, 550Ω Max. 
For Voltage o/p, 3000Ω Min. 

OUTPUT ACCURACY 

±0.25% of span 

* With Analog Output, CE marking is not applicable/valid.

 

 

Programming and Setting

 

KEYPAD 

8-keys tactile membrane 
keypad  

CONFIGURATION 
SOFTWARE 

All Configurable parameters can 
be set through PC Based 
software 

MEMORY 

Non-volatile, restored after 
power loss 

Communication Specification

 

NO. OF COMMUNICATION 
PORT 

2-RS485(COM-1 and 
COM-2) . COM2 is 
Optional 

COMMUNICATION TYPE 

Half duplex/Asynchronous 

COMMUNICATION 
PROTOCOL 

MODBUS RTU. All 
parameters are 
Configurable through 
MODBUS Protocol. 

MAXIMUM NO. OF UNITS 

32 

COMMUN. ERROR 
DETECTION  

CRC Check 

 

PROFIBUS Communication (Optional) *

 

MODE 

Profibus DP Slave 

BAUD RATE 

9600, 19.2K, 44.45K, 93.75K, 
187.5K, 500K, 1.5M, 12M bps 

ADDRESS 

Configurable through Configuration 
Software (0 to 125 Only) 

NETWORK 
CAPACITY 

 

Multi-drop up to 31 modules, Plus 
a host, without a repeater 

 

Up to 125 modules plus a host if 
four repeaters are used  

COMMUICATION 
DISTANCE 

Up to 1200 meters without a 
repeater using Type A wire  

 

1200m @ 115Kbps or less 

 

1000m @ 187.5Kbps 

 

400m @ 500Kbps 

 

200m @ 1.5Mbps 

 

100m @ 12Mbps 

* With Profibus communi. , CE marking is not applicable

 

 

HMI Interface (Optional)*

 

NO. OF COMMUNICATION 
PORT 

1-RS-232 (HMI) 

COMMUNICATION TYPE 

Half 
duplex/Asynchronous 

COMMUNICATION 
PROTOCOL 

MODBUS RTU  

CONNECTABLE NO. OF 
UNITS 

COMMUN. ERROR 
DETECTION  

CRC Check 

 

Network Connectivity (Optional)

 

NO. OF COMMUNICATION 
PORT 

1(RJ-45) 

TRANSMISSION SPEED 

10 Mbps 

NETWORK PROTOCOL 

TCP/IP 

APPLICATION PROTOCOL 

MODNET 

 

Data logging (Optional)

 

Data logging Memory Type 

Flash Memory (32 
MB) 

Data logging type 

Periodic and Event 

Periodic Memory Size 

25 MB 

Event Memory Size 

7 MB 

RTC Time format 

DD/MM/YY – 
HH:MM:SS 

Periodic Logging sampling 
time 

1 Second minimum 

Event polling time 

1 second  

USB Port* 

1(USB 2.0) 

USB Function  

For retrieving logged 
data only 

Max. USB storage device 
size 

Upto 32 GB 

USB Mass storage device 
format 

 

FAT16 

 

FAT32 

USB fetched data file format 

.xls (only) 

USB data retrieving option 

 

Full Data Fetch  

 

Fetch Data by time 

*With USB port, CE marking is not applicable/valid.

 

 

Display Specification

 

CHANNEL NO 
DISPLAY 

2-digits, 7-segment, Green , 0.56” 
character height 

DATA 
DISPLAY 

4-digits, 7-segment, Red, 0.56’’ 
character height 

PARAMETER 
DISPLAY 

6-digits, 16-segment Alphanumeric, 
Orange LEDs, 0.3” character height 

STATUS LEDs 

24-Red  LEDs  for  Alarm,  24-Orange 
LEDs Control Output, 8-Green LEDs for 
Relay, 1-Red LED for Manual mode, 1- 
Green  for  Run  mode,1-Red  for  Fault, 
2-Green(Rx) 

2-Red(Tx) 

for 

Communication 

 

Environmental Specification:

 

AMBIENT TEMPERATURE 

-10 to 55°C 

HUMIDITY 

30% to 95% RH 
(Non-Condensing) 

TEMPERATURE COEFFICIENT 

< 100ppm 

WEIGHT 

1.25 KG 

INSTRUMENT WARM-UP 
TIME 

<15 mins after power 
on 

DEGREE OF PROTECTION 

IP54 (From Front) 

 

Power Supply Specification

 

RATED VOLTAGE 

85-265VAC-50/60Hz / 
100-300VDC or 18-36VDC 

POWER CONSUMPTION 

Max. 16 VA (85-265 VAC) 
and Max. 8 VA (18-36 
VDC) 

 

Isolations (Withstanding Voltage) 

 

 

Between primary terminals* and secondary 

terminals**: 1500VAC for 1 minute

 

 

 Between secondary terminals: 500V AC for 1 

minute 
* Primary terminals indicate power terminals and 
relay   output terminals 
** Secondary terminals indicate analog input signals, 
Digital Contact output terminals, communication  
terminals and Ethernet N/W terminal 

 

Insulation Resistance: 

20MΩ or more at 500 V DC  

 

Signal Isolation Specifications

 

Sr.No  Signals 

Signal Isolation 

Power Input 

Isolated  from  other  ip/op 
terminals and internal circuit 

Analog Inputs 

Not isolated from other analog 
i/p  terminals  &  the  internal 
circuit. But isolated from other 
ip/op terminals. 

RS-485 
Communication 

Isolated  from  other  ip/op 
terminals and internal circuit 

Ethernet 
Communication 

Isolated  from  other  ip/op 
terminals and internal circuit 

Relay contacts 

Isolated  between  contact  o/p 
terminals  &  from  other  ip/op 
terminals and internal circuit 

Digital Output 

Isolated  from  other  ip/op 
terminals and internal circuit 

                  Quick User Guide

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

85XX+ Scanner/DAQ

 

NUMBER OF 
INPUT 
CHANNELS 

16 

RATED 
INPUT 
VOLTAGE  

12 V DC    
(Sink / 
Source) 

24 V DC    
(Sink / 
Source) 

110 V 
DC 
(Sink) 

220 V 
DC  
(Sink) 

INPUT ON 
VDC 

≥ 7 V  

≥ 15 V  

≥ 75 V   ≥ 110 V  

INPUT OFF 
VDC 

≤ 4 V  

≤ 5 V  

≤ 30 V   ≤ 50 V  

INPUT 
CURRENT  

4 mA ± 
20% / 
Ch 

4 mA ± 
20% / 
Ch 

2 mA / 
Ch 

2 mA / 
Ch 

MAXIMUM 
ALLOWABLE 
INPUT 
VOLTAGE 

15 V DC 

30 V DC 

132 V 
DC 

250 V 
DC 

RESPONSE 
TIME 

50 mSec 