Calibration Mode for Technical Maintenance 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TABLE OF CONTENTS 

 

Introduction …………………………………………………………2 

 

Power supply ……………………………………………………… 2 

 

Low battery ………………………………………………………… 2 

 

Specifications ……………………………………………………… 2 

 

Display and key function ………………………………………..    3 

 

Calibration ………………………………………………………….. 4 

 

Trouble error measures ………………………………………..5 & 6 

 

 

 

 

 

 

 

 

 

 

 

 

 

START 

Press Test pin twice which at rear of the scale to 

enter Test mode. 

Press ON/OFF/ZERO key to zero the scale. 

Display shows span count 

(4000~~14000).

 

Display shows CAL    

 

Press ON/OFF/ZERO key once. 

Display shows calibration 

weight in normal mode. 

     End 

Load calibration weight 