QUI-1046 TAG Read Kit Installation
1
TubeTAG
PLUS TM
Installation Manual
Version 1.2 December 2009
Note: Operators instructions can be found on the Technical CD supplied with the U
LTR
A (
50:50
)
Page 1: ...QUI 1046 TAG Read Kit Installation 1 TubeTAGPLUS TM Installation Manual Version 1 2 December 2009 Note Operators instructions can be found on the Technical CD supplied with the ULTRA 50 50 ...
Page 2: ...TRA 50 50 Remove the covers from ULTRA 50 50 Loosen the 2 bolts holding the oven securing plate in place Raise the securing plate and retighten the 2 bolts ULTRA 50 50 only Loosen the white thumb screw then tilt the fan backwards Loosen the 2 screws highlighted ...
Page 3: ...ad Kit Installation 3 Tilt the tube oven assembly towards the rear of the instrument Feed the connector and wire from the TAG reader through the opening highlighted Feed the TAG reader plate through the opening shown ...
Page 4: ...the plate on the 2 loosened screws The heat and sensor plate needs to be replaced on system with serial no before GB00 M30764 To remove the plate first disconnect the sensor by carefully pulling out the connector The plate can then be lifted out ...
Page 5: ...ensuring it sits flat in the groove Attach connector to the sensor Tilt the tube oven back in place and secure the TAG reader plate On ULTRA 50 50 with serial no GB00 M2xxxx Mount the TAG PCB on the rear left of the instrument as shown Note Ensure board does not contact the control PCB ...
Page 6: ...LTRA 50 50 with serial no GB00 M3xxxx Mount the TAG PCB on the rear left of the instrument on the mounting plate highlighted Plug the ribbon cable and TAG reader connector into the PCB Secure any loose wires with cable ties Replace all covers ...