background image

GAUGE PLATE 
BRACKET

INTERMEDIATE 

PLATE

The G1109 90° peel fi xture mounts to a Mark-10 ESM303 test stand. The illus-
tration at left depicts proper installation. Please follow these instructions:

1.      Mount the intermediate plate to the base plate of the ESM303 test stand 

with four socket head screws.

2.     Mount the fi xture onto the intermediate plate with four socket head screws.

3.     Install a force gauge onto the force gauge mounting plate using only the 

upper two thumb screws, provided with the ESM303. 

4.     Install the G1078 grip onto the gauge plate bracket. 

5.      Attach  the  gauge  plate  bracket  to  the  back  of  the  force  gauge  mounting 

plate by installing the two provided thumb screws through the correspond-
ing clearance holes, into the threaded holes in the force gauge housing. 

6.    Position the crosshead at an appropriate starting position for the test.

7.      Pull the loose end of the pull cable from the rear of the fi xture until the work 

table has reached the end of its leftward travel.

8.        

Consider asking for the assistance of a colleague with this step. 

While 

holding the work table in place at its end of travel, wrap the pull cable at 
least three times around the bollard on the G1078 grip, then route the cable 
through the slot and tighten with the thumb screw.  

9.      If the optional AC1080 grip adapter is required, install it to the work table 

with four socket head screws. 

10.  

This step is not necessary, but is recommended to protect the pull cable.

Connect the limit switch cable from the connector on the right side to the 
connector  at  the  rear  of  the  ESM303  labeled  “AUX.  LIMITS”.  When  the 
work table reaches the end of travel, the test stand will stop and the the 
indicator in front of the fi xture will illuminate. 

If using the optional AC1078 or AC1079 adjustable angle adapter, refer to the 
installation instructions on Page 2. 

ESM303 

TEST STAND

WORK TABLE

G1078 GRIP

FORCE GAUGE 

MOUNTING PLATE

G1109 90° Peel Fixture - Installation Instructions

Mark-10 Corporation

  

  

www.mark-10.com

  

  

[email protected]

Toll-free:  888-MARK-TEN

  

  

Tel: 631-842-9200

  

  

Fax: 631-842-9201        

 

 

     

32-1232

AC1080 GRIP 
ADAPTER 
(OPTIONAL)

LIMIT 

SWITCH 

CABLE 

CONNECTOR

PULL CABLE

CROSSHEAD

PAGE 1 OF 2

Reviews: