MESSAGES
9.1
MESSAGES DURING SELF-TEST
9.2
ERROR MESSAGES
9.2.1
IN SPEED DISPLAY
9.2.2
IN SPEED AND FLOW DISPLAYS
Speed
display
Flow
display
Status
display
Meaning
EX.X
[X.X]
: Software version
RFC XXXX
[XXXX]
: Operating hours since the ROTAFLOW
Console’s last servicing
RFDMXXXX [XXXX]
: Operating hours since the master
ROTAFLOW Drive’s last servicing
RFDS– – – –
Slave ROTAFLOW Drive not connected
–OK–
EX.X
PUMP OK
Self-test completed without errors.
Message
Acoustic
alarm
Pump
stop
Cause/Error source
⇐⇐
-
-
Recommendation to turn flow regulator to zero.
⇑⇑⇑⇑
Yes
No
Speed > upper speed limit (HLIM)
⇓⇓⇓⇓
Yes
No
Speed < lower speed limit (LLIM)
!!!!
Yes
Yes
More than one
[RUN AWAY]
within 20 seconds
Switch off pump to clear error.
Message
Acoustic
alarm
Pump
stop
Cause/Error source
ERR OR
-
-
Error: The status display shows a further message.
LOW BAT.
Yes
No
The batteries are almost depleted.
voltage during battery operation”, page 40
NO SEL
Yes
Yes
Console control module in the heart-lung machine