Message
Possible conse
quences
Possible cause
Possible action
Pat.: Water tempera
ture too high
Temperature at the out
let too high. If error is
not rectified: the device
heats up further until a
safety shutdown is initi
ated. The device can
then no longer be used
for a period of time.
Overshoots generated
by regulator
Reduce setpoint tem
perature in order to
avoid overshoots. If the
fault still recurs, notify
authorized service.
Mains power fail
No circulation and thus
no temperature regula
tion possible.
Mains power has failed Restore mains supply.
Confirm the error mes
sage on the CU. Switch
off active circuits. Ma
chine is ready for use
again (internal UPS
buffers control unit for
10 mins). Use replace
ment device.
Safety pump stop!
Valve calibration
(~100s)! Start pump af
ter valve calibration.
Valve calibration
Wait 100 seconds.
6.4.2 Medium Priority
Message
Possible conse
quences
Possible cause
Possible action
Cplg.: Water flow too
low!
No circulation and thus
no temperature regula
tion possible.
Outlet valve closed.
Open outlet valve.
Tube kinked.
Remove kink.
Cplg.: No gradient
mode possible...
Gradient mode not
available.
External temperature
sensor defective or
connection faulty.
Check connection to
external temperature
sensor. Reconnect if
necessary.
Cplg.: Water pressure
too high!
No circulation and thus
no temperature regula
tion possible.
Outlet valve closed.
Open outlet valve.
Tube kinked.
Remove kink.
Cplg.: Tank tempera
ture too high!
If the water tempera
ture exceeds 42°C:
The safety shutdown is
triggered. No circulation
until the water has
cooled down.
Temperature in the
tank too high or tem
perature sensor defec
tive
Activate ice formation.
Replace water in tank
with colder water. If the
error occurs again: No
tify authorized service.
Use replacement de
vice.