| HCU 40 |
11 Technical Data
| 137 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
Permissible pressure
The HCU 40 can generate a maximum pressure
of 2 bar.
For oxygenator heat exchangers and cardiople
gia heat exchangers with lower permissible max
imum pressures, a lower pressure limit of the
HCU 40 must be set.
Oxygenator heat exchangers with a maximum
permissible pressure of < 1.5 bar can be used
with a HCU 40 in 60 Hz mains operation in the
cardioplegia water circuit.
11.3.2 Blanket Connection Kit (incl. Pressure Reducer)
Nominal pressure of pressure reducer
Max. 1.2 bar
11.4 Components Supplied
HCU 40
Power cable
Instructions for Use
Control unit CU (incl. holder and pole for securing)
CAN connection cable for HCU 40 – CU connection, length 1 m
Tubing set for oxygenator/heat exchanger with 1/2" Hansen coupling (straight or angled), 2 x 6 m
Tubing set for cardioplegia heat exchanger with 3/8" Hansen coupling (straight or angled), 2 x 6 m
2 cleaning connectors (double Hansen connectors)
Drainage tube
Bypass tubing set for patient water circuit with 1/2" Hansen coupling, length 1 m
Bypass tubing set for cardioplegia water circuit with 3/8" Hansen coupling, length 1 m
Disinfectant, powder, 500 g (active ingredient: Tosylchloramide sodium (Chloramine T))
Mixing bottle for diluting the disinfectant
11.5 Ambient Conditions
Operation
Storage and transport
Temperature
+10 … +30ºC
20 … +60ºC
Relative humidity (noncon
densing)
0 … 85%
10 … 96%
Air pressure (absolute)
800 … 1060 hPa
700 … 1060 hPa