7 Cleaning, descaling and disinfection
| 125 |
Instructions for Use | 1.0 | EN | 01
Copyright Maquet Cardiopulmonary GmbH
7.2.4 Emptying the Tank
This function empties the tank. You must use this function to change the water.
If you only want to empty the tubes of a circuit instead, you can use the "Empty
tubes" function (⇨ "Emptying Water Circuits", page 58).
The HCU 40 supports you during emptying with a wizard to guide you step by
step.
1
Touch the
[Settings]
symbol.
2
Touch the
[Cleaning/emptying]
symbol.
3
Touch the
[Emptying the tank]
symbol.
4
You will see a wizard that will guide you
through the emptying process.
5
Disconnect all heat exchangers which are connected to the patient water
circuit and/or the cardioplegia water circuit, and connect the supplied cleaning
connectors instead.
6
Touch the
[Next]
symbol and wait until the melting process is complete.
▶
The HCU 40 melts the ice block and adjusts the temperature of the tank
water to 20°C. Duration: 30 up to a maximum of 75 minutes (depending
on the amount of ice present and the power supply).