Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
55
Technical specifications
9
Electromagnetic compatibility (EMC)
9.6
Electromagnetic compatibility (EMC)
The produkt ist designed for operation in the
environment specified below. The customer
has to ensure that the product is operated in an
environment meeting these specifications.
9.6.1
Electromagnetic emission
9.6.2
Electromagnetic interference resistance
Measurement /
Standard
Compliance
Electromagnetic environment / Guidelines
RF radiation
CISPR 11
Group 1
The product uses RF energy exclusively for its own
internal functions. Consequently its RF emission
levels are very low and it is improbable that it will
cause interference in nearby electronic equipment
Class B
The product is intended for use in all facilities, to in-
clude residential areas, and in those facilities which
are connected directly to a public power supply net-
work which also serves residential buildings.
Harmonics
IEC 61000-3-2
Class A
Voltage fluctuations /
Flicker
IEC 61000-3-3
Complies
Fig. 42: Electromagnetic emission
Measurement /
Standard
Test level
Compliance level
Electromagnetic envi-
ronment / Guidlines
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 6 kV
Contact discharge
± 8 kV
Air gap discharge
± 6 kV
Contact discharge
± 8 kV
Air gap discharge
Floors should be made of
wood or concrete or be
finished with ceramic
tiles. If the floor is covered
with a synthetic material,
then the relative humidity
shall be at least 30 %.
Rapid electrical
transients /
Bursts
IEC 61000-4-4
± 2 kV
for mains cords
± 1 kV
for input and output ca-
bles
± 2 kV
for mains cords
± 1 kV
for input and output ca-
bles
The quality of the mains
power voltage should cor-
respond to that of a typi-
cal commercial or
hospital environment.
Surges
(surges)
IEC 61000-4-5
± 1 kV
Differential mode voltage
± 2 kV
Common mode voltage
± 1 kV
Differential mode voltage
± 2 kV
Common mode voltage
The quality of the mains
power voltage should cor-
respond to that of a typi-
cal commercial or
hospital environment.
Fig. 43: Electromagnetic interference resistance (Part 1 of 2)