Technical specifications
Technical specifications
9.4
Technical specifications
9.5
Vacuum in dependency on altitude
Nominal voltage
230 V, AC power
Rated frequency
50 Hz
Power consumption
92 W
Max. power consumption
700 mA
Fuse
2 x T 1.25 A/H; 250 V; 5 x 20 mm
Operating mode
Intermittent operation:
30 min. on / 45 min. off
Displacement, product at sea level
16 ± 3 l/min.
Max. vacuum at sea level
approx. -80 kPa
Vacuum regulation
Shunt air screw, mechanical
Altitude
Pump ultimate vacuum Pump ultimate vacuum
2000 m
-58 kPa
-435 mm Hg
1500 m
-63 kPa
-473 mm Hg
1000 m
-69 kPa
-518 mm Hg
500 m
-74 kPa
-555 mm Hg
0 m
-80 kPa
-600 mm Hg