1133.12B1/B3/F1/F3
GA113312EN02
8
III. General description
1.
General features
In princliple, the operating table 1133.12 ALPHAMAXX
is suitable for the following surgical procedures:
• in the thorax and abdominal areas
• in the head area
• in gynaecology
• in urology
• in orthopaedics
If required, separate accessories can be attached to
the operating table.
The table top is subdivided into 6 sections:
• head plate (removable)
• upper back plate (removable)
• lower back plate
• seat plate
• split leg plates
The lower back, seat and leg plates are adjusted under
motor power.
The head plate and the upper back plate are adjusted
manually.
X-ray cassettes can be inserted underneath the back
plate.
The side rails are used to attach accessories.
For patients weighing up to 135 kg (300 lb), place
the operating table on its castors (UNLOCK) only
up to the maximum table top height of 940 mm
(37 in)! The warning sign is not seen!
For patients weighing between 135 kg and max.
225 kg (300 - 500 lb), place the operating table
on its castors (UNLOCK) only when the table top
is in its lowest position.
Using the UNLOCK function for patient weights
exceeding 225 kg (500 lb) is not permissible.
1.1 Versions
Various versions of the operating table 1133.12 are avail-
able.
1133.12B1
Operating table incl. SFC padding and
standard side rails
1133.12B3
Operating table incl. SFC padding, stand-
ard side rails and autodrive in the base
1133.12F1
Operating table incl. SFC padding and US
side rails
1133.12F3
Operating table incl. SFC padding, US side
rails and autodrive in the base
The corded hand control, the IR transmitter, the charg-
ing station for the IR transmitter as well as head plates,
upper back plates and leg plates are to be ordered sepa-
rately. The order numbers are listed in chapters VIII and
IX.
NO
YES
Fig.III-1:
The preferred patient orientation
III. General description
Fig.III-2:
Warning sign for table top height