17
System Reset
The system can be reset at any time.
To reset system manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Diagnostics
and press
ENTER
.
3.
In Diagnostics menu, arrow down to select
System Reset
and press
ENTER
.
Set PIN Number
The user can set a PIN number for added security for the operational settings of the L100 analyser. Once this
function is enabled, other users will not be able to change the parameter settings without the PIN number but
will still be able to run analysis or access data logs.
To set PIN number manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Diagnostics
and press
ENTER
.
3.
In Diagnostics menu, arrow down to select
Set PIN Number
and press
ENTER
.
4.
Use the Up and Down arrow touch keys to set the value.
Note:
Default value is set to 0000 and security is disabled.
Erase Logs
To erase logs manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Diagnostics
and press
ENTER
.
3.
In Diagnostics menu, arrow down to select
Erase Logs
and press
ENTER
.
4.
In the Erase Log menu, select:
a)
Erase Result Log
(to clear the result log), or
b)
Erase Event Log
(to clear the event log), or
c)
Erase Cal. Log
(to clear the calibration log) and press
ENTER
.
Note that it is recommended that the logs be deleted every time the sensor is changed to prevent data build up
and slow log response. The logs can be downloaded as a .csv file from the PeCOD® prior to deletion for record
keeping. Logs are downloaded to a PC using the PeCOD® software supplied by MANTECH.