4.8 Other Operational Functions
Event Log
Please refer to section 6, Troubleshooting
Run Standby
This function can be activated when the L100 is not in use and ensures that the instrument is primed and
maintains calibration for the next sample, when left inactive for several hours during the day. When this function
is enabled, the L100 analyzer will prime Port B.
To set “Periodic Standby” manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Set Up
and press
ENTER
.
3.
In the Set Up menu, arrow down to select
Periodic Actions
and press
ENTER
.
4.
In the Periodic Actions menu, select
Standby Interval
and press
ENTER
.
5.
Use the Up, Down, Left and Right arrow touch keys to set the value for standby interval time in hours:
minutes format.
6.
Press
ENTER
.
To run Standby manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Operation
and press
ENTER
.
3.
In Operation menu, arrow down to select
Run Standby
and press
ENTER
.
The analyzer will perform the run standby function periodically according to the interval time set in step (5),
starting from the time the function was activated.
4.9 Diagnostic Functions
Factory Settings
The operational values can be reset back to default (factory) settings at any time.
To reset operational values back to factory settings manually from the touchpad:
1.
Press
MENU
.
2.
Arrow down to select
Diagnostics
and press
ENTER
.
3.
In Diagnostics menu, select
Factory Settings
and press
ENTER
.