Other tasks
Chapter 4
Visco 88
Page 4-5
To change the current set of measuring constants:
1.
Turn off the instrument.
2.
Turn the
SYSTEM
switch to position 2.
3.
Turn the
SPEED
switch to the set of measuring systems to use. (Position 1
selects set 1, position 2 selects set 2, etc.).
4.
Press the
START
button and hold it down.
5.
Switch on the power. The LCD will read '
----
' at first, then after two seconds
display
t
ES
t
.
6.
Release the
START
button. The display will show which set of constants is
selected (
Se
t.n
where n is
1
to
8
) for a couple of seconds and will then go into
normal measuring mode. For example, the second set would be displayed as
Se
t.
2
.
7.
The set selection is stored while the instrument is powered off. The current set
is indicated by a small number from 1 to 8 at the bottom of the display.
Freezing the display
The last measured value can be 'frozen' by pressing the
CAL
button. To do this,
the motor must be running and the viscometer must be under front panel control
(not software control).
In
Freeze mode
the mode indicator flashes. Once the current measurement has
been frozen, the display can be stepped as usual with the
FUNCTION
button to
display any of the measuring points at the time when the
CAL
button was pressed.
To return to the normal display updates, press
CAL
again. (Freeze mode is also
terminated if any of the
SPEED
or
SYSTEM
selectors are changed or the motor is
stopped).
Note
Do not press
CAL
immediately after changing the speed (i.e. when the
LCD reads '----') as a few seconds are necessary for Freeze mode to
become active.
i
Summary of Contents for Visco 88
Page 1: ...Visco 88 User Manual MAN0378 Issue 1 0 October 2006...
Page 6: ...Table of Contents Visco 88 Page iv MAN 0378...
Page 7: ...Part 1 Operator s Guide...
Page 8: ......
Page 22: ...Chapter 2 What is the Visco 88 Page 2 12 MAN 0378...
Page 34: ...Chapter 3 Making a measurement Page 3 12 MAN 0378...
Page 40: ...Chapter 4 Other tasks Page 4 6 MAN 0378...
Page 54: ...Chapter 6 Maintenance and troubleshooting Page 6 6 MAN 0378...
Page 55: ...Part 2 Appendices...
Page 56: ......
Page 62: ...Appendix A Health and safety Page A 6 MAN 0378...
Page 69: ...Visco 88 Page D 1 D Regulatory statements EMC...
Page 70: ...Appendix D Regulatory statements Page D 2 MAN 0378 D...