Mai ntenance
60
Step
Action
should be felt when it is seated.
Note
: Do not apply any additional pressure.
3
Remove the tool.
Install the syringe glass
To install the syringe glass, follow the instruction video and/or the software guided work-
flow integrated in the MicroCal PEAQ-ITC Control Software.
The software guided workflow is also shown in the following table.
Step
Action
1
Grasp the pipette's rotating assembly.
2
Gently insert the syringe glass with the syringe's fill port facing the same direction as the
hole in the pipette's rotating assembly.