
H ealth and safety
14
!
Warning! Hazardous substances and biological agents
. When using haz-
ardous chemical and biological agents, take all suitable protective measures,
such as wearing protective glasses and gloves resistant to the substances used.
Follow local and/or national regulations for safe operation and maintenance of
MicroCal PEAQ-ITC.
Installing and moving
!
Warning!
The product must be installed and prepared by Malvern Instruments
personnel or a third party authorized by Malvern Instruments.
!
Warning! Power cord
. Only use power cords with approved plugs delivered or
approved by Malvern Instruments.
!
Warning! Electrical power
. Provide proper electrical power to the instrument.
This should be 100-240 VAC, 50/60 Hz, observing any local electrical safety
requirements. Only power cords supplied with the instrument should be used.
!
Warning! Protective ground
. The product must always be connected to a
grounded power outlet.
!
Warning! Access to power switch and power cord with plug
. Do not block
access to the power switch and power cord. The power switch must always be
easy to access. The power cord with plug must always be easy to disconnect.
!
Warning!
Do not block the ventilation inlets or outlets on the system.
!
Notice! Disconnect power.
To prevent equipment damage, always disconnect
the power from the product before an instrument module is removed or
installed, or a cable is connected or disconnected.