30
|
Motors do not
operate via the
hand controller
Mains cable is
not plugged in
Plug in the
mains cable
Socket not live
Check socket
Cabling plug
connection not
sitting correctly
Check plug
connection
to motor and
hand controller
Hand controller
or motor is
faulty
Inform the
operator, re-
tailer or Betten
Malsch GmbH
Locking the
functions on
the hand con-
troller
Releasing
the functions
on the hand
controller
Adjustments
using the elec-
tric motor are
not functioning
properly
There is an
obstacle in the
adjustment
area
Check moving
parts and
remove any
foreign bodies
The safe work-
ing load has
been exceeded
Reduce the
load
Motors cut out
after a longer pe-
riod of operation
The adjustment
time or safe
working load has
been exceeded,
and the control
unit has reacted to
overheating
Allow the motor
system to cool
down properly
Opposite function
activates when
operating the
hand controller
buttons
Motor plugs are
mixed up
Check the correct
cables or inform
your operator,
retailer or Betten
Malsch GmbH
The side rails can
no longer be set
up properly
The mechanism is
obstructed or bent
Check all moving
parts and remove
any foreign bod-
ies, or contact our
customer service
team
Castors do not
brake or do not
roll
A foreign body
has got into the
rollers
Remove foreign
bodies
The castor system
is faulty
Contact our
customer service
team
T R O U B L E S H O O T I N G