2.9
Commissioning (To only be completed by qualified persons)
a)
Before switching on the device make sure the
DEVICE IS LEVEL
and
WATER TAPS ARE ON
. Check that the
DRAIN WASTE
is connected.
b)
Turn on the power at the isolation switch and press the standby button on the front display. The digital display will
illuminate.
c)
Check that the water tank is filling with cold water and that it fills to Level 3. Ensure the lid is put back on water tank.
d)
Check that the steam generator tank has filled to the full Level 3. The steam generator will preheat to 85°C in standby
mode or during cycle.
e)
Check that the 5-litre detergent bottle has been fitted.
f)
Flush approximately ½ litre of water down the steam generator tank overflow pipe. This will fill the ‘S’ Trap at the hose
junction and prevent steam coming back up into the water tank.
g)
Open the door and turn power off at the control. Check that flush nozzle rotate freely. Check that all spray nozzles are
tight. Turn power back on and close door by pressing the manual door button.
h)
Re-install all covers
i)
The device will display ‘rdY’ (ready) when the water levels are reached.
j)
Run a cycle and conform cycle is complete.
Note:
DO NOT USE THE DEVICE WITHOUT THE WATER SUPPLY TURNED ON