malema.com
M-100
Adjustable flow switch
with right-angle flow
Operating Instructions
and
Quick Start Guide
Page 1: ...malema com M 100 Adjustable flow switch with right angle flow Operating Instructions and Quick Start Guide ...
Page 2: ...alema com Table of Contents Introduction 03 Operation 03 Storage and Handling 03 Installation Instruction 03 Specification 05 Certification 06 Dimensional drawings 07 Ordering Information 07 Warranty 08 ...
Page 3: ...firm that the model code on the label matches the specification in the purchase order Installation Instructions This product is mounted vertically models suitable for Universal mounting are available on request Adequate filtration and sealing procedures should be used when mounting in flow line To ensure protection of malema flow switches from moisture penetration of lead wires a suitable connecto...
Page 4: ...s bulletin it is advisable to keep within those ratings as excessive power will FRY the switch Electrical Protection Ratings The electrical Reed Switch is rated for several million cycles and should rarely require replacement Most switch failures are the result of overloading DO NOT EXCEED RATED CAPACITY as this will cook the REED SWITCH Catalogue ratings are based upon D C resistive loads however...
Page 5: ...al Ratings Reed Switch Switching Voltage Breakdown Voltage DC Resistive AC Resistive Switching Current 10 Watts SPST or 3 Watts SPDT Hermetically Sealed UL Recognized File E47258 Operating Temperature 40o C to 125o C 200 VDC 170 VDC for SPDT 250 VDC 200 VDC for SPDT 10 Watts 3 Watts for SPDT 10 VA 3VA for SPDT 0 5 A 0 25 A for SPDT 1 2 A 0 5A for SPDT LAMP DERATED BY 2 3 Lead Wires No 24 to 18 AWG...
Page 6: ... meets CE compliance under LVD 2014 35 EU RoHS and REACH can be obtained from Malema on request UL Recognised 1 UL and Canadian UL Recognized for use in Class I Groups A B C and Class II Groups E F G all divisions hazardous locations Details can be found on http www malema com certifications 2 Certified for ATEX and IECEX directives Details can be found on http www malema com certifications 3 Cert...
Page 7: ...00 Model with 1 4 ports Illustrated is the M 100 Model with 1 8 ports and universal mounting Standard Part Numbering Options M Model Material Port Switch Flow Range Mounting Piston Seals M 100 S 1 1 1 0 0 0 100 B Brass S 316SS T PTFE 1 1 8 2 1 4 1 SPST N O 2 SPST N C 3 SPDT 1 Low 2 High 0 Standard Vertical 1 Universal Mounting with spring 0 Standard 316SS with epoxy 1 PTFE encapsulated 2 All 316SS...
Page 8: ... to the buyer Service and Repair To return the products please obtain an RMA number for the product by contacting Malema Sensors Corporate Office Boca Raton at 800 637 6418 or 561 995 0595 All returns of equipment must go to the following address Malema Sensors 1060 S Rogers Circle Boca Raton FL 33487 USA NOTE Specifications are subject to change without notice Warranty 08 2018 Malema Engineering ...