6.3. Measuring Light
•
Set the function switch to the “Lux” scale and set the range to
desired (“Lux” or “x10 Lux”) range.
•
Move the meter to face the photo detector to a light source in a
horizontal position.
•
Read the illuminance nominal from the LCD display.
•
Over-range: If the instrument only display one “1” in the M.S.D. the
input signal is too strong, and a higher range should be selected.
•
When the measurement is completed. Remove the photo detector
from the light source.
•
Spectral sensitivity characteristic: To the detector, the applied photo
diode with filters makes the spectral sensitivity characteristic almost
meet C.I.E. (International Commission on Illumination) photopia
curve V (
λ) as the following chart described.
100% (Relative Sensitivity) Spectral Sensitivity
Wave length (nm)
80%
60%
40%
20%
0
V(
λ
)
400 500 600 700
10
Recommended Illumination:
Locations
Lux
•
Office
Conference, Reception room
200 to 750
Clerical work
700 to 1,500
Typing drafting
1000 to 2,000
•
Factory
Packing work, Entrance passage
150 to 300
Visual work at production line
300 to 750
Inspection work
750 to 1,500
Electronic parts assembly line
1500 to 3,000
•
Hotel
Public room, Cloakroom
100 to 200
Reception, Cashier
200 to 1,000