
Operation Manual MB 11 BERAphone
®
/Classic
8100496 Rev. 6
78
23/10/2019
Calibration values
RadioEar IP30 with Coupler IEC 60318-4 (IEC 711):
TRANSDUCER
CE-CHIRP
®
pe RETSPL [dB re. 20 µPa]
Radioear IP30 with eartips
30
Radioear IP30 with EarCups
TM
56.5
RadioEar IP30 with Coupler IEC 60318-5:
TRANSDUCER
CE-CHIRP
®
pe RETSPL [dB re. 20 µPa]
Radioear IP30 with eartips
20.5
Radioear IP30 with EarCups
TM
47
The calibration values for eartips are according to PTB report from 2008-05-19, in
compliance to the calibration procedure defined in standard EN 60645-3. Correction
values for EarCup
TM
calibration values are defined as MAICO standard values.
BERAphone
®
with coupler 60318-3:
TRANSDUCER
CE-CHIRP®
pe RETSPL [dB re. 20 µPa]
BERAphone
®
32.5
The calibration values for the BERAphone
®
are according to the PTB Report from
2008-05-19. The values are equal to the click stimulus calibration values for the MB 11
BERAphone
®
regarding the standard ISO 389-6. Both stimuli have the identical
amplitude magnitude spectrum. A special calibration adapter is needed to mount the
BERAphone
®
onto the coupler with a pressure of 5 N.
Sound attenuation characteristics
Frequency[Hz]
Sound Attenuation [dB] ISO 4869-1
RadioEar IP30
125
33.5
250
34.5
500
34.5
1000
35.0
2000
33.0
4000
39.5
8000
43.5
Summary of Contents for MB 11 BERAphone
Page 1: ...Operation Manual MB 11 BERAphone Classic ...
Page 2: ......