13 Appendix A: EMC Compatibility
Portable and Mobile RF communications equipment can affect
the MAICO MA 27/MA 27e. Install and operate the MAICO MA
27 / MA 27e according to the EMC information presented on this
page and the next 5 pages.
The MAICO MA 27 / MA 27e has been tested for EMC emissions
and immunity as a standalone instrument. Do not use the
MAICO MA 27 / MA 27e adjacent to or stacked with other
electronic equipment. If adjacent or stacked use is necessary,
the user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those
specified, with the exception of servicing parts sold by MAICO
as replacement parts for internal components, may result in
increased EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for
making sure the system complies with the IEC 60601-1-2
standard.
EUT Support Equipment and Cables*
Item
Manufactur
er
Model
Cable
SIP/SOP
Serial
no.
Length
[meter]
Screened
[Y/N]
Socket
ID
Type
Audiometric
Headset
Radioear
DD45
2.0
Y
Phones
L and R
Headset
output
-
Patient
response
switch
DGS
APS3
2.9
Y
Patient
Resp.
DC
level
-
Power
Supply
Fuhua
UE24WCP-
050250SPA
1.5
N
USB/DC-
in
DC
level
-
*To ensure compliance with the EMC requirements as specified
in IEC 60601-1-2, it is essential to use only the supplied support
equipment and cables from MAICO.