36
BA315001-en
9.13
Adjusting the IR Test Bench (MET 6.1 / 6.3 only)
DANGER
•
Carbon monoxide (CO) is a colourless and odourless gas and therefore not
perceivable.
•
At the end of a gas calibration the bottle with calibration gas must always be
closed.
•
At the first sign of symptoms indicating carbon monoxide poisoning start first
aid treatment.
Symptoms and First Aid with Carbon Monoxide Poisoning
•
Headaches, dizziness, ear buzzing and near collapse (after several hours or
towards the end of the working day) are symptoms of CO positioning.
•
Smoking increases the CO-concentration in blood and can lead to poisoning
symptoms even when there is a low CO content in the air.
•
Acute life threatening condition if fainting or unconsciousness occurs!
(the person has red face coloring
•
Close off the calibration gas bottle immediately.
•
Provide an oxygen mask for the sick person!
•
Contact emergency medical services immediately!
•
Air out the rooms!
For the adjustment, you will need:
•
Calibration gas with the following composition:
CO: 1.2 to 8.2 vol. %; CO
2
: 7.5 to 16 vol. %; C
3
H
8
: 550 to 12 000 ppm
•
Pressure regulator (part # 28000092)
•
Connection tubing suitable for 5 mm barbed fitting
•
MAHA Emission Software
Procedure:
1 Start the MET service tool via:
MES: Otto > Device diagnosis > Adjustment
or Windows: Start Programs > Maschinenbau Haldenwang > EUROSYSTEM
X.XX > MET service tool
2 Click the "Gas adjustment" button to begin adjustment.
3 Enter the gas components of your calibration gas in the input mask.
4 Click the "1-point adjustment" button to start the adjustment process.
5 Connect the calibration gas cylinder to the MET's calibration gas inlet. If the
pressure regulator illustrated is used, open the valve fully. For other designs,
set a pressure of 0.5 to 1 bar and click OK in the information dialogue to
confirm.