 Connecting the measuring unit

Th

  is measuring unit uses an interface that is compliant with the 

USB (Universal Serial Bus) 2.0 FS (Full speed) standard, and can be 
connected directly to a personal computer or hub.

Note

Use a personal computer that is compatible with the recommended 
system requirement (refer to “Specifi cations”).

 Installing software

Th

 e soft ware available on the Magnescale website can be used by 

installing in a personal computer. Refer to the soft ware instruction 
manuals available on the Magnescale website for the soft ware 
installation and operation methods.

Magnescale website
https://www.magnescale.com/mgs/language/english/product/
Digital Gauge > DS805S/DS812S

 Operating Cautions

•  Do not use the unit in places where it may receive excessive 

shocks. Otherwise the inside of the unit may be damaged or the 
unit may become unable to produce normal output signals.

•  Be sure to turn off  the power before connecting or disconnecting 

connectors in order to prevent damage or misoperation.

• Aft er mounting the measuring unit, insert the connector into the 

display unit until it locks. Be sure to turn off  the power switch 
before connecting or disconnecting the connector.

•  Locate the measuring unit (head built-in portion) at least 10 cm 

away from a strong magnetic source. (Do not exceed 5 mT in 
parallel magnetic fi elds.)

•  Do not forcibly pull the cable for connecting or disconnecting, or 

it may cause breakage.

•  Th

  e measuring unit is short circuited to the frame GND (ground 

terminal) of the display unit by shield wire of the cable. When 
the measuring unit is used with an other signal processing device 
on a machine tool, etc., be aware of the ground level.

•  Th

  e N-type measuring unit has an unsealed spindle for use in 

mechanical devices and measuring rooms having a favorable 
work environment. 

  Due to the low measurement force, the N-type cannot be used 

with the spindle directed upward.

•  Th

  e P-type measuring unit is dust-proof, drip-proof, and oil-

resistant making it suitable for use in environments subject to 
chips, cutting oil, and machine oils. However, avoid immersing 
the P-type in cutting oil, etc., as it is not of a water-proof 
structure.

•  Adhesion of cutting oil, etc. which increases in viscosity aft er 

drying, may hinder accurate spindle motion. Remove them with 
water or tepid water. Th

  en apply machine oil or the like to the 

spindle.

•  Th

  e P-type measuring unit is provided on the sliding portion of 

the spindle with an anti-drip/dust seal of nitrile rubber. Spindle 
movement may be impaired by exposure to cutting oil, water, etc. 
that may remove oil from the spindle surface.

•  Recommended calibration interval 1 year.

 Mounting Instructions

•  Be sure to chuck the stem for mounting the measuring unit. If 

other portion of the measuring unit needs to be chucked, avoid 
chucking the name label portion (where the head is built in) and 
its neighborhood (Fig. 1).

 

If the measuring unit is mounted on another device, ensure that 
the measuring unit is free from contact with mount screws.

•  If excessive torque is applied on the spindle when screwing in 

the feeler, the internal mechanisms can be damaged. Th

 erefore, 

screw in the feeler with your hands, and never use a tool.

  Except when using a fl at feeler, it is recommended to either 

attach the spring washer (nominal size: 2.5) or use a screw lock 
to prevent loosening of the feeler. (Tightening torque reference 
value: 0.05–0.06 N·m)

•  Th

  e measuring unit comprises ball bearings. Th

 erefore, chucking 

the stem of the measuring unit too tightly when mounting may 
damage the spindle and prevent its smooth motion.

•  A holder of dimensions as shown in Fig. 2 may be used to mount 

the measuring unit, tightening the screw to about 4 N·m torque.

•  Since mounting parallelism aff ects the measuring accuracy, 

adjust the parallelism to within 0.3 mm in relation to 100 mm 
displacement by means of mount bracket (or sleeve).

•  Fix the cable in a suitable position to prevent possible cable 

breakage. Never handle the cable by forcibly pulling or bending 
it. (Inside bend radius 50 mm or more)

•  If the spindle is to be fi xed to the machine, use coupling DZ-191 

which is optionally available. Refer to the DZ-191 Instruction 
Manual regarding the mounting procedure.

• Aft er the stem is secured, do not apply force in the rotational 

direction. Th

  is could cause a failure.

20H6

M4

30

2

25

40

24

8

Chucking portions

Sleeve

Fig. 1  Mounting positions

φ

4.5, 

φ

8 Spot facing depth 4

Tightening torque: 4 N

m

Use 4M hexagon socket head cap screw.

Fig. 2  Mount holder dimensions

Unit: mm

φ

 +0.013

      0

Head built-in portion

Stem

 Troubleshooting

Th

  e operating status of the measuring unit can be checked using the LED 

indicator.

LED lighting color

Details

Remedial action

Lights in blue.

The power is turned on. Normal operation 
is now performed. The reference point has 
been detected.

-

Blinks in blue.

The power is turned on.
The reference point standby status is 
established or the reference point is being 
detected.

When the reference point is used, turn on the power with the spindle 
extended as far as possible (Note: in case of SV type, with the spindle 
contracted), and then move the spindle 1.5 mm or more.
When the power is turned on with the spindle stopped near the reference 
point position, it may not be possible to detect the reference point or the 
reference point may shift.

Blinks in red.

The signal level of the measuring unit is 
abnormal.

After dealing with the factor suspected to be responsible for the alarm, reset 
the software or turn the power back on.
If the same symptoms persist even when the power has been turned back on, 
the measuring unit might be malfunctioned.

Lights in red.

The maximum response speed has been 
exceeded. The signal level of the measuring 
unit is abnormal. Serial communication 
trouble has occurred. Normal operation 
cannot be performed due to the eff ects 
such as external noise etc.

Off 

Power OFF

If the same symptoms persist even when the power has been turned on, the 
measuring unit might be malfunctioned or the cable might be broken.

LED indicator

 Dimensional Diagrams

*

 When cable extension is needed, please use CE38 which is optionally available. (Cable length Max. 4 m)

DS50NR5 / PR5

DS100NR5 / PR5

66
70

18

15

286

12

(102)

329.5

443.5

(20)

2.5

38.5

(20)

(51)

12

44

223

1

Pin arrangement Signal name

2
3
4

VBUS

D-

D+

GND

φ

6

 0 − 0.013

φ

20

− 0.02 − 0.07

φ

20

USB2.0TYPE-A Connector

Interpolation box

(St

em)

C

able length 2 m

Cable length
0.5 m 

(St

em)

Unit: mm

 0 − 0.013

φ

20

φ

8

φ

4.8

φ

4.8

φ

25

 0 − 0.1