12.1.2
Running a test sequence
Starting a sequence
1. Select the desired sequence from the Immittance module control panel.
2. Click
Start
to start the selected sequence.
Warning
•
If the patient is troubled by the test, stop the test. The test is interrupted immediately.
Already measured results are kept.
In reflex testing the data collection will be paused before each stimulus above 100 dB HL, and you will
be prompted to continue.
Stopping a sequence
If a leak is detected or the probe is removed from the patient's ear, the sequence stops automatically.
1. Click
Stop
to stop the sequence manually.
2. If needed, perform supplementary measurements. You can supplement an interrupted sequence by
deselecting the successful part and then restart the sequence testing.
12.2
Quick Check testing
You can perform Tympanometry and Reflex Screening as a Quick Check test from the
Tymp + Reflex Screening
screen.
Note
•
The safety intensity level cannot be exceeded in screening mode.
Preparing for testing
1. Do as described in
.
2. Launch the Immittance module with the
Tymp + Reflex Screening
screen selected.
3. Ask the patient to sit very still and quiet during the test, without moving head or jaw.
4. If needed, toggle the
Ear Selector
to select the ear on which you wish to start the test.
5. Load a test setting configured for Quick Check testing of patients in the relevant age group.
Procedure
1. To prepare the probe for Quick Check testing, press the
Start
button on the probe or shoulder pad before you insert
it into the ear canal of the patient.
Otometrics - MADSEN Zodiac
87
12 Testing with Zodiac Diagnostic or Clinical - PC-based
Summary of Contents for zodiac
Page 1: ...MADSEN Zodiac Reference Manual Doc No 7 50 1610 EN 02 Part No 7 50 16100 EN...
Page 21: ...4 2 Bottom view A Probe connections B Label Otometrics MADSEN Zodiac 21 4 Views of Zodiac...
Page 156: ...jB jB App 2 5 2 G App 2 5 3 B G B G Y B G B G 156 Otometrics MADSEN Zodiac App 2 MADSEN Zodiac...
Page 157: ...App 2 5 4 226 Hz 226 Hz 200 daPa Otometrics MADSEN Zodiac 157 App 2 MADSEN Zodiac...
Page 206: ...152 151 151 152 150 152 206 Otometrics MADSEN Zodiac Index...