MadgeTech VFC2000-MT Product User Manual Download Page 2

 Product User Guide  |  2

VFC2000-MT

PRODUCT USER GUIDE

Product Overview

The VFC2000-MT is a simple solution for vaccine 
temperature monitoring compliance. Designed to satisfy 
all CDC and VFC requirements, the VFC2000-MT provides 
precise, continuous temperature monitoring and validation 
for temperatures as low as -100 °C (-148 °F). Featuring a 
convenient LCD screen, the VFC2000-MT displays current 
readings, minimum and maximum statistics as well as a 
battery level indicator. User-programmable alarms trigger 
an audible and visual alert. Optional glycol bottle monitors 
for temperatures as low as -50 °C (-58 °F) and AC power 
source allows battery to be a back up if there is a power loss. 

VFC Requirements

•  Detachable, buffered temperature probe
•  Out-of-range audible and visual alarms
•  Low battery indicator with external power and battery 

backup

•  Current, minimum, and maximum temperature display
•  Accuracy of ±0.5°C (±1.0°F)
•  Programmable logging interval (1 reading per second 

to 1 reading per day)

•  Daily check reminder warning
•  Suitable for vaccine transportation
•  Also monitors ambient room temperature

Device Operation

1.  Install the MadgeTech 4 Software onto a Windows PC.
2.  Connect the data logger to the Windows PC with the 

provided USB cable.

3.  Launch the MadgeTech 4 Software. The VFC2000-MT  

will appear in the Connected Devices window 
indicating the device has been recognized.

4.  Select the start method, reading interval and any other 

parameters appropriate for the desired data logging 
application. Once configured, deploy the data logger by 
clicking the Start icon.

5.  To download data, select the device in the list, click the 

Stop icon, and then click the Download icon. A graph 
will automatically display the data.

Selection Buttons

The VFC2000-MT is designed with three selection buttons:

Scroll:

 Allows user to scroll through 

current readings, average stats, daily 
minimum and maximum temperatures 
and device status information displayed on 
the LCD screen.

°C

°F

Units: 

Allows users to change displayed 

units of measurement to either Celsius or 
Fahrenheit.

Start/Stop: 

To activate Manual Start, 

arm the device through the MadgeTech 4 
Software. Hold the button for 3 seconds. 
There will be two beeps confirming the 
device is started.  Reading will be displayed 
on the screen and status in software will 
change from 

Waiting to Start

 to 

Running

To pause logging while running, hold 
button for 3 seconds. 

LED Indicators

Status:

 Green LED blinks every 5 seconds 

to indicate the device is logging.

Check:

 

Blue LED blinks every 30 seconds to 

indicate daily statistic check has gone past 
24 hours. Hold Scroll button for 3 seconds to 
reset reminder.

Alarm:

 

Red LED blinks every 1 second to 

indicate an alarm condition is set. 