PRTrans1000IS

PRODUCT USER GUIDE

Product Overview

The PRTrans1000IS samples up to 100 Hz (10 ms) but 
records data based on user-defined pressure thresholds. 
The device can be set to start the trigger on high and/or 
low thresholds and end the trigger after either a chosen 
number of samples, or when the signal crosses over 
the set stop threshold. The device can record up to 255 
events and take as many as 262,143 pressure samples. 
The PRTrans1000IS has been certified as intrinsically safe 
certification in accordance with the latest issue of FM3600 
and FM3610.

Certified Intrinsically Safe for:

•  Class 1 Division 1 Group ABCD
•  Class 1 Division 2 Group ABCD
•  Temperature Class: T4A  

Operational Warnings

•  When used in hazardous locations, the PRTrans1000IS is 

to be 

installed prior

 to the location becoming hazardous 

and removed only after the area is no longer hazardous.

•  The maximum allowed ambient temperature for the 

PRTrans1000IS (under any circumstances) is 80 °C. The 
minimum rated operating temperature is -40 °C.

•  The PRTrans1000IS is approved for use only with the 

Tadiran TL-2150/P battery. Replacement with any other 
battery will void the safety rating.

•  Batteries are user replaceable, but are to be removed or 

replaced only in locations known to be non-hazardous.

•  Tampering or replacement of non-factory components 

may adversely affect the safe use of the product, and is 
prohibited. Except for replacement of the battery, the 
user may not service the PRTrans1000IS. MadgeTech, 
Inc. or an authorized representative must perform all 
other service to the product.

Installation Guide

Installing the Software

The Software can be downloaded from the MadgeTech 
website at 

madgetech.com

. Follow the instructions 

provided in the Installation Wizard.

Installing the Interface Cable

To access the COM Port for the interface cable, unscrew the 
key-ring end cap. Connect the device into a USB port with 
the interface cable. The drivers will install automatically. 
Screw the end cap back onto the data logger until the 
O-ring cannot be seen, before deploying the data logger.

Pressure Sensor

To use the pressure sensor for gauge measurements, 
screw the ¼ inch male NPT fitting into the pipe to be 
measured with a 9⁄16 inch wrench. 

Ordering Information

•  901271-00  — PRTrans1000IS-1000-PSIA
•  901275-00 — PRTrans1000IS-100-PSIA
•  901279-00 — PRTrans1000IS-100-PSIG
•  901283-00 — PRTrans1000IS-300-PSIA
•  901287-00 — PRTrans1000IS-300-PSIG
•  901291-00 — PRTrans1000IS-30-PSIA
•  901295-00 — PRTrans1000IS-30-PSIG
•  901299-00 — PRTrans1000IS-5000-PSIA
•  901303-00 — PRTrans1000IS-500-PSIA
•  901307-00 — PRTrans1000IS-500-PSIG
•  900298-00 — IFC200 USB Interface Cable
•  901744-00 — Tadiran TL-2150/P Replacement Battery