Device Maintenance
Device Operation (cont’d)
5. Screw the cover onto the data logger.
Downloading Data from a Data Logger
1. Place the logger into the docking station (IFC400 or
IFC406).
2. Highlight the data logger in the
Connected Devices
list. Click
Stop
on the menu bar.
3. Once the data logger is stopped, with the logger
highlighted, click
Download
.
4. Downloading will offload and save all the recorded
data to the PC.
Battery Replacement
Materials:
Replacement Battery (Tadiran TL-2150/S)
1. Move device to a non-hazardous location before
replacing battery.
2. Observe Operational Warnings when removing and
replacing the battery.
3. Unscrew the bottom of the data logger and remove the
battery.
4. Place the new battery into the logger. Caution: Observe
To ensure desired operation of the TEMP1000Ex,
please keep the surface clear of any foreign objects or
substances
. The TEMP1000Ex’s data is downloaded through external contact with the IFC400 or IFC406 docking
station. Covering the surface with foreign objects
(i.e. Calibration Labels)
can prevent the communication and/
or downloading process.
O-ring maintenance is a key factor when properly
caring for the Temp1000Ex. The O-rings ensure a tight
seal and prevent liquid from entering the inside of the
device. Please refer to the application note “O-Rings 101:
Protecting Your Data”, found at
, for
information on how to prevent O-ring failure.
Recalibration
Recalibration is recommended annually. To send devices
back for calibration, visit
Additional Services:
Custom calibration and verification point options
available, please call for pricing.
correct battery polarity when installing.
Temp1000Ex
PRODUCT USER GUIDE
Communication
O-Rings
Call for custom calibration options to accommodate specific application needs.
Prices and specifications subject to change. See MadgeTech’s terms and conditions
at madgetech.com.
To send devices to MadgeTech for calibration, service or repair, please use the
MadgeTech RMA Process by visiting madgetech.com.