background image

HiTemp140-M12

6

Temperature Sensor

Various RTD Probes with M12 Connector 

(Sold Separately)

Measurement Range

18 

Ω

 to 400 

Ω

-200 °C to +850 °C (-328 °F to +1562 °F) 

(Probe Dependent)

Temperature Resolution

0.0001 

Ω

0.01 °C (0.018 °F) 

(Probe Dependent)

Calibrated Accuracy

±0.015 

Ω

±0.05 °C (±0.09 °F) 

(Probe Dependent) 

*Calibrated 

Accuracy Range

18 

Ω

 to 200 

Ω

 

-200 °C to +265 °C (-328 °F to +509 °F)

Connector Type

M12 Female, 4 Pin

Reading Rate

4 readings per second up to 1 reading every 24 hours

Memory

43,690 readings

Start Modes

•  Software programmable immediate start
•  Delay start up to 18 months in advance

Stop Modes

Manual or Timed (specific date and time)

Trigger Settings

High & Low limits may be set. Once data meets or exceed sets limits, the device will 
record to memory. Bi-level start and stop triggers can also be programmed. Users can 
specify the number of readings to take after the device triggers. 

Readings in Trigger  

Settings Mode

18,724 readings

Real Time Recording

May be used with PC to monitor and record data in real time

Password Protection

An optional password may be programmed into the device to restrict access  
to configuration options. Data may be read out without the password.

Memory Wrap Around

Yes

Battery Type

3.6V high-temperature lithium battery included; 

user replaceable

Battery Life

1 year typical (1 minute reading rate)

Calibration

Digital calibration through software

Calibration Date

Automatically recorded within device

Data Format

Date and time stamped °C, °F, °R, K

Time Accuracy

±1 minute/month at 25 ºC

Computer Interface

IFC400 or IFC406 USB docking station required; 125,000 baud

Operating System  

Compatibility

XP SP3/Vista/Windows 7/Windows 8

MadgeTech Software  

Compatibility

•  MadgeTech Standard Software version 4.2.5.0
•  MadgeTech Secure Software version 4.2.4.0 or later

Operating Environment

-40 ºC to +140 ºC (-40 ºF to +284 ºF), 0 %RH to 100 %RH non-condensing,  
0.002 PSIA to 60 PSIA

IP Rating

IP40 - logger alone (with no probe connected). The logger will inherit IP rating of the 
attached probe (up to, and including IP68/IP69K).

Dimensions 

(body)

2.15 in x 0.97 in x 0.97 in (54.5 mm x 24.6 mm x 24.6 mm)

Dimensions 

(probe)

Probe dependent

Weight

85 g (3.0 oz)

Materials

Body: 316 Stainless Steel

Approvals

CE

HiTemp140-M12 General Specifications

*Calibrated accuracies based on standard MadgeTech calibrations for 0 Ω to 200 Ω range.

Summary of Contents for HiTemp140-M12

Page 1: ...To view the full MadgeTech product line visit our website at www madgetech com Product User Guide HiTemp140 M12 HiTemp140 M12 High Temperature Data Logger with an M12 Probe Connector ...

Page 2: ...line visit our website at www madgetech com HiTemp140 M12 Data Logger Product Overview 2 Installation Guide 3 Device Operation 3 Wiring the Data Logger 4 Probe Connection 4 Product Maintenance 5 RMA Instructions 5 General Specifications 6 Table of Contents ...

Page 3: ...ote Specific M12 wiring configuration is required for compatibility see page 4 Submergibility The HiTemp140 M12 logger itself is rated IP40 and not submersible It will inherit the IP rating of the probe it is attached to allowing for potential submergibility depending upon the probe O Rings O ring maintenance is a key factor when properly caring for the HiTemp140 M12 The O rings ensure a tight sea...

Page 4: ...e for the data logging application and click Start Quick Start applies the most recent custom start options Batch Start is used for managing multiple loggers at once Real Time Start stores the dataset as it records while connected to the logger The status of the device will change to Running Waiting to Start or Waiting to Manual Start depending upon your start method Disconnect the data logger fro...

Page 5: ... attached to up to and including IP68 IP69K Probe Connection Connecting probes to the HiTemp140 M12 Materials 15 mm wrench Compatible Probe with M12 Connector 1 Connect the probe to the data logger using the wiring diagram above 2 Use a 15 mm hex wrench to tighten the threads and secure the probe to the data logger The HiTemp140 M12 is compatible with probes that have the wiring configuration as s...

Page 6: ...ce signed in click on the Make New RMA button and fill in all the blank fields 3 Complete the applicable fields on the form including customer Billing and Shipping information even if they are the same Please see the field explanation below for a more detailed description about questions asked in the Device Information section 4 When all of the fields are complete click Generate RMA 5 Print out th...

Page 7: ...ammed into the device to restrict access to configuration options Data may be read out without the password Memory Wrap Around Yes Battery Type 3 6V high temperature lithium battery included user replaceable Battery Life 1 year typical 1 minute reading rate Calibration Digital calibration through software Calibration Date Automatically recorded within device Data Format Date and time stamped C F R...

Page 8: ...MadgeTech Inc 6 Warner Road l Warner NH 03278 Phone 603 456 2011 l Fax 603 456 2012 www madgetech com l info madgetech com DOC 1312036 00 REV 2 01 22 2018 ...

Reviews: