Device Operation (cont’d)
Trigger Settings
The device can be programmed to only record based off
user configured trigger settings.
1. In the
Connected Devices
panel, click the device desired.
2. On the
Device
Tab, in the
Information
Group, click
Properties
. Or, right-click the device and select
Properties
in the context menu.
3. Select
Trigger
in the Properties window.
4. Trigger formats are available in
Window
or
Two Point
Mode
. Window mode allows a high and/or low trigger set
point, and a trigger sample count or “window” of time
recorded when set points are exceeded to be defined. Two
point allows for different Start and Stop setpoints to be
defined for both the high and low triggers.
Refer to the
Trigger Settings - MadgeTech 4 Data Logger
Software
for instructions on
how to configure Trigger Settings.
Set Password
To password protect the device so that others cannot
start, stop or reset the device:
1. In the
Connected Devices
panel, click the device
desired.
2. On the
Device
Tab, in the
Information
Group, click
Properties
. Or, right-click the device and select
Properties
in the context menu.
3. On the
General
Tab, click
Set Password
.
4. Enter and confirm the password in the box that
appears, then select
OK
.
Device Maintenance
O-Rings
O-ring maintenance is a key factor when properly
caring for the HiTemp140-FP. The O-rings ensure a tight
seal and prevent liquid from entering the inside of the
device. Please refer to the application note
O-Rings 101:
Protecting Your Data
, found at
, for
information on how to prevent O-ring failure.
Battery Replacement
Materials:
ER14250MR-145 Battery
1. Unscrew the bottom of the logger and remove the
battery.
2. Place the new battery into the logger. Note the polarity
of the battery. It is important to insert the battery with
positive polarity pointing upward towards the probe.
Failure to do so could result in product inoperability or
potential explosion if exposed to high temperatures.
3. Screw the cover back onto the logger.
Recalibration
MadgeTech recommends annual recalibration. To send
devices back for calibration, visit
.
Note:
This product is rated for use up to 140 °C
(284 °F). Please heed the battery warning. The
product will explode if exposed to temperatures
above 140 °C (284 °F).