Temperature Sensor

36 inch Glass Braided RTD depyrogenation probe

Temperature Resolution

0.0001 

Ω

, 0.01 °C (0.02 °F)

Temperature Accuracy

±1.2 °C (-200 °C to +400 °C)

Probe Measurement Range

-200 °C to +400 °C

Connector Type

M12 Female, 4 Pin

Reading Rate

4 readings per second up to 1 reading every 24 hours

Memory

43,690 readings

Start Modes

•  Software programmable immediate start
•  Delay start up to 18 months in advance

Stop Modes

Manual or Timed (specific date and time)

Trigger Settings

High & Low limits may be set. Once data meets or exceed sets limits, the device 
will record to memory. Bi-level start and stop triggers can also be programmed. 
Users can specify the number of readings to take after the device triggers. 

Readings in Trigger  

Settings Mode

18,724 readings

Real Time Recording

May be used with PC to monitor and record data in real time

Password Protection

An optional password may be programmed into the device to restrict access  
to configuration options. Data may be read out without the password.

Memory Wrap Around

Yes

Battery Type

3.6V high-temperature lithium battery included; 

user replaceable

Battery Life

1 year typical (1 minute reading rate)

Calibration

Digital calibration through software

Calibration Date

Automatically recorded within device

Data Format

Date and time stamped °C, °F, °R, K

Time Accuracy

±1 minute/month at 25 ºC

Computer Interface

IFC400 or IFC406 USB docking station required; 125,000 baud

Operating System Compatibility

XP SP3/Vista/Windows 7/Windows 8/Windows 10

Dimensions

Body: 2.15 in x 0.97 in x 0.97 in (54.5 mm x 24.6 mm x 24.6 mm)
Probe: 0.10 in dia. x 36 in L (2.54 mm dia. x 76.2 mm L)

Materials

Body: 316 Stainless Steel 
Probe: Stainless Steel, Fiberglass, PEEK

Approvals

CE

Battery Warning

WARNING: FIRE, EXPLOSION, AND SEVERE BURN HAZARD. DO NOT SHORT CIRCUIT, CHARGE, FORCE OVER 
DISCHARGE, CRUSH, PENETRATE, OR INCINERATE. BATTERY MAY LEAK OR EXPLODE IF HEATED ABOVE 150ºC (302ºF).

General Specifications

Product User Guide

Summary of Contents for DHS

Page 1: ... Red 485 Dark Blue Pantone 289 Light Blue 58 of Pantone 289 To view the full MadgeTech product line visit our website at www madgetech com DHS Data Logging System Dry Heat Sterilization Data Logging System Product User Guide ...

Page 2: ...ighlight the desired data logger For most applications select Custom Start from the menu bar and choose the desired start method reading rate and other parameters appropriate for the data logging application and click Start Quick Start applies the most recent custom start options Batch Start is used for managing multiple loggers at once Real Time Start stores the dataset as it records while connec...

Page 3: ...l Barrier 1 Unscrew the lid of the thermal barrier 3 Unscrew the HiTemp140 M12 data logger from the lid of the thermal barrier 4 Slide the probe of the data logger out from the lid of the thermal barrier 2 Slide out the data logger and lid from the barrier body ...

Page 4: ... back in to MadgeTech follow the instructions below to create an RMA Return Merchandise Authorization on the MadgeTech website 1 Visit www madgetech com under the Services tab select RMA Process 2 When the web page opens please sign in If this is the first time select Click here to register an account and create one Once signed in click on the Make New RMA button and fill in all the blank fields 3...

Page 5: ...password may be programmed into the device to restrict access to configuration options Data may be read out without the password Memory Wrap Around Yes Battery Type 3 6V high temperature lithium battery included user replaceable Battery Life 1 year typical 1 minute reading rate Calibration Digital calibration through software Calibration Date Automatically recorded within device Data Format Date a...

Page 6: ...ech com Operating Environment Refer to the Time vs Temperature Chart IP Rating Not Rated Barrier Materials 316 Stainless Steel PTFE Silicone Enclosure Dimensions 1 75 in OD x 9 6 in L 44 45 mm OD x 243 8 mm L Weight 3 0 lb 1350 g Thermal Barrier Specifications DHS Data Logging System ...

Page 7: ... 601 minutes 160 C 468 minutes 170 C 396 minutes 180 C 348 minutes 190 C 313 minutes 200 C 286 minutes 210 C 265 minutes 220 C 247 minutes 230 C 233 minutes 240 C 220 minutes 250 C 209 minutes 260 C 200 minutes 270 C 192 minutes 280 C 184 minutes 290 C 178 minutes 300 C 172 minutes 310 C 166 minutes 320 C 161 minutes 330 C 157 minutes 340 C 153 minutes 350 C 400 C 60 minutes Product User Guide Tim...

Page 8: ...o potentially unsafe levels The ThermoVault Max may take hours to fully cool down Even if the exterior of the ThermoVault Max is cool to the touch the interior of the barrier and its contents may still be VERY hot The ThermoVault Max is primarily intended for use in dry air environments but with the addition of the TMAX Wet Seal Kit the ThermoVault Max may also be used in liquids and steam environ...