Product User Guide | 2
CryoTemp
PRODUCT USER GUIDE
Product Overview
The CryoTemp data logger is designed for use in ultra-
low operating temperatures for processes such as
monitoring blood plasma, vaccines, pharmaceuticals,
frozen foods and shipping containers. The CryoTemp
can record temperatures as low as -86 °C (-122 °F). This
standalone device does not require any additional probes.
The enclosure is designed with a built-in handle for easy
attachment and is IP64 splash resistant.
Installation Guide
Installing the Software
The Software can be downloaded from the MadgeTech
website at
. Follow the instructions
provided in the Installation Wizard.
Installing the Docking Station
IFC300 (sold separately)
— Connect the device into a USB
port with the interface cable and install the drivers.
Ordering Information
• 900038-00 — CryoTemp
• 900315-00 — IFC300
Device Operation
Connecting and Starting the Data Logger
1. Once the software and USB drivers are installed, plug
the USB cable into the docking station.
2. Connect the other end of the USB cable into an open
USB port on the computer.
3. Place the data logger into the docking station.
4. The data logger will automatically appear under
Connected Devices
within the software.
5. For most applications, select
Custom Start
from
the menu bar and choose the desired start method,
reading rate and other parameters appropriate for the
data logging application and click
Start
.
(
Quick Start
applies the most recent custom start options,
Batch
Start
is used for managing multiple loggers at once,
Real Time Start
stores the dataset as it records while
connected to the logger.)
6. The status of the device will change to
Running
,
Waiting to Start
or
Waiting to Manual Start
,
depending upon your start method.
7. Disconnect the data logger from the interface cable
and place it in the environment to measure.
Note:
The device will stop recording data when the end of memory is
reached or the device is stopped, unless user selectable memory wrap
is enabled. At this point the device cannot be restarted until it has been
re-armed by the computer.
Downloading Data from a Data Logger
1. Connect the data logger to the computer with the
interface cable.
2. Highlight the data logger in the
Connected Device
s
list. Click
Stop
on the menu bar.
3. Once the data logger is stopped, with the logger
highlighted, click
Download
.
4. Downloading will offload and save all the recorded
data to the PC.