MadgeTech CryoTemp Product User Manual Download Page 2

 Product User Guide  |  2

CryoTemp

PRODUCT USER GUIDE

Product Overview

The CryoTemp data logger is designed for use in ultra-
low operating temperatures for processes such as 
monitoring blood plasma, vaccines, pharmaceuticals, 
frozen foods and shipping containers. The CryoTemp 
can record temperatures as low as -86 °C (-122 °F). This 
standalone device does not require any additional probes. 
The enclosure is designed with a built-in handle for easy 
attachment and is IP64 splash resistant.

Installation Guide

Installing the Software

The Software can be downloaded from the MadgeTech 
website at 

madgetech.com

. Follow the instructions 

provided in the Installation Wizard.

Installing the Docking Station

IFC300 (sold separately)

 

— Connect the device into a USB 

port with the interface cable and install the drivers.

Ordering Information

•  900038-00 — CryoTemp
•  900315-00 — IFC300

Device Operation

Connecting and Starting the Data Logger

1.  Once the software and USB drivers are installed, plug 

the USB cable into the docking station.

2.  Connect the other end of the USB cable into an open 

USB port on the computer.

3.  Place the data logger into the docking station.
4.  The data logger will automatically appear under 

Connected Devices

 within the software.

5.  For most applications, select 

Custom Start

 from 

the menu bar and choose the desired start method, 
reading rate and other parameters appropriate for the 
data logging application and click 

Start

(

Quick Start

 

applies the most recent custom start options, 

Batch 

Start

 is used for managing multiple loggers at once, 

Real Time Start

 stores the dataset as it records while 

connected to the logger.)

6.  The status of the device will change to 

Running

,  

Waiting to Start

 or 

Waiting to Manual Start

depending upon your start method.

7.  Disconnect the data logger from the interface cable 

and place it in the environment to measure.

Note:

 The device will stop recording data when the end of memory is 

reached or the device is stopped, unless user selectable memory wrap 
is enabled. At this point the device cannot be restarted until it has been 
re-armed by the computer.

Downloading Data from a Data Logger

1.  Connect the data logger to the computer with the 

interface cable.

2.  Highlight the data logger in the 

Connected Device

s

 

list. Click 

Stop

 on the menu bar.

3.  Once the data logger is stopped, with the logger 

highlighted, click 

Download

4.  Downloading will offload and save all the recorded 

data to the PC.