Instruction for use URYXXON
®
500
11
Instruction for use URYXXON
®
500 V 2.1 / 01.23
4. Initial operation
This chapter describes the first steps to a successful start‑up.
The correct execution of the steps described here must always be
ensured by the user.
4.1 Insert the transport rack
NOTICE
Only start the instrument once the transport rack has been
inserted and all power connections are complete.
Before turning on the instrument for the first time, the transport
rack has to be inserted. Remove the black transport tray
h
from
the instrument. To do so, pull on the black handle on the transport
tray’s right side.
Before you insert the transport rack, make sure there is no
packaging material blocking the mechanics. Take the transport
rack
b
and align its feed holes with the corresponding rubber
pins on the transport panel.
The mounting direction of the transport rack is determined by the
asymmetrical feed holes / pins.
Push the transport tray gently back into the instrument.
4.2 Power supply
CAUTION
Multiple hazards: Only qualified personnel should perform the
instructions described in this step.
WARNING
Risk of electric shock. Only use the included power
supply Type SM‑075 A0F‑11. Use of other power
supplies can damage the device and harm the user.
CAUTION
Risk of injury due to the danger of tipping over due to slipping.
Place the device on a flat surface. Do not stack the device.
WARNING
Take care not to overload the mains socket. There is a risk
of overload and fire. Make sure that the power cord is not
damaged. Check the suitability of the power source used for the
equipment. Only connect the power supply to grounded outlets.
WARNING
Risk of injury: Familiarize yourself with the device before working
with it and read this document carefully. Do not use the device
unless you have received instruction in its use.
NOTICE
Do not switch on the device until all data and power cables have
been fully assembled and connected.
The instrument requires a power supply of 7.5 V DC, which is
supplied by the power pack ((Type SM‑075 A0F‑11) 100 – 250 V,
47 · 63 Hz) or GST40A07 / Globtek Inc GTM96600‑4507.5T3.