Installation with a mounting plate
DO NOT cut the sensor cable. Re-termination of the sensor is highly specialised and MUST be carried out at
a MACE technical facility.
A mounting plate is normally used in large pipes where a cost-effective solution is required. The sensor is
attached to the plate, which is then bolted to the inside surface of the pipe. This plate is supplied in
polypropylene.
1. Attach the sensor head securely to the mounting plate by the six 3mm screws supplied (M3 x 8 mm
Phillips head countersunk 304 s/s).
WARNING: Use only a hand screwdriver. Do not over tighten these screws as
permanent damage to the thread may result
2. Tie the cable to the mounting plate using the holes and cable ties provided. Make sure that the knob
of the cable tie sits to the back of the assembly, out of the flow.
DO NOT OVERTIGHTEN CABLE
TIES.
3. The cable from the sensor to the XCi device should then be run inside the pipe and secured to the pipe
in such a manner as to prevent damage by debris in the pipe. The cable should be routed through
electrical conduit.
MACE XCi User Manual - 48 -
Summary of Contents for HVFlo XCi
Page 1: ......
Page 10: ...MACE XCi User Manual 10...
Page 18: ...MACE XCi User Manual 18...
Page 21: ...MACE XCi User Manual 21...
Page 50: ...MACE XCi User Manual 50...
Page 98: ...MACE XCi User Manual 98...
Page 173: ...MACE XCi User Manual 173...
Page 178: ...MACE XCi User Manual 178...
Page 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Page 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Page 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Page 250: ...MACE XCi User Manual 250...