My meter isn't reading the right flow rate!
Symptom
Cause
Remedy
My meter is over reading!
Incorrect pipe diameter entered
in device configuration
Incorrect pipe diameter units
(
US ONLY
)
Check you've not mixed up "ft"
with "in" or
vice versa
Pipe has silt in the bottom
Check/clear silt from pipe
Velocity range not set correctly
Ensure
set to
Sensor installed too close to
undershot gate
Check
and rectify
My meter is under reading!
Incorrect pipe diameter entered
in device configuration
Incorrect pipe diameter units
(
US ONLY
)
Check you've not mixed up "ft"
with "in" or
vice versa
Insert sensor not installed far
enough into pipe
Check
and rec-
tify
Insert sensor not installed per-
pendicular to flow
Check
and
rectify
My meter is reading neg-
ative flows!
Sensor is installed facing down-
stream
of
sensor
Sensor is installed facing down-
stream
Change
in
Velocity range not set correctly
Ensure
set to
Leaking gate/foot valve
Check seals of gates and valves
My meter is NOT reading
flow when it IS flowing!
XCi device not "
Started
"
device using
Doppler sensor is covered with
trash
Check sensor for trash and clear
MACE XCi User Manual - 220 -
Summary of Contents for HVFlo XCi
Page 1: ......
Page 10: ...MACE XCi User Manual 10...
Page 18: ...MACE XCi User Manual 18...
Page 21: ...MACE XCi User Manual 21...
Page 50: ...MACE XCi User Manual 50...
Page 98: ...MACE XCi User Manual 98...
Page 173: ...MACE XCi User Manual 173...
Page 178: ...MACE XCi User Manual 178...
Page 211: ...For a short video of the procedure click Play MACE XCi User Manual 211...
Page 238: ...3 Unscrew the locking ring using the tool and remove completely MACE XCi User Manual 238...
Page 239: ...4 Remove the damaged ceramic depth sensor MACE XCi User Manual 239...
Page 250: ...MACE XCi User Manual 250...