47DAC OPERATING MANUAL EM-9518-B Rev.4
24.2 OPERATING PROCEDURE
NOTE
The following figures are display examples. The displays
depend on the settings.
1
Hold down Alarm/
↓
, Scale/
↑
and Shift buttons at once for 5
seconds or more to move on to Loop Test Output Mode.
• The measuring is stopped and the last measured values or status
are held for the DC, alarm and BCD outputs.
• The current indication starts blinking, to which you can apply
changes.
• The SD indicates ‘TEST UP’.
NOTE
• In moving on to Loop Test Output Mode while ‘S.ERR’ is indicated, 110% of the display scaling span is indicated when
the SD indicates ‘OVER’.
• In moving on to Loop Test Output Mode while ‘-20000’ blinks, -10% of the display scaling span is indicated.
• In moving on to Loop Test Output Mode while ‘100000’ blinks, 110% of the display scaling span is indicated.
• The specified decimal point position is applied.
1,3
1,3
1,2
2
Blinking