Maintenance
Maintaining the vacuum part
26
Vacuum Regulator for Chlorine Gas
C 2700-V
© Lutz-Jesco GmbH 2016
Operating Instructions
BA-21210-02-V05
10.5.3 Installing the vacuum part
Precondition for action:
The parts are completely dry following cleaning.
The parts are in good condition.
Wearing parts are available.
Perform the following working steps:
1.
Rub a little silicone grease onto the edges of the diaphragm (
).
2.
Put the diaphragm (
) with the curvature facing downward onto
the diaphragm disc (
) and then attach the clamping ring (
3.
Firstly only put the screws (
) and some fitting grease loosely into
place.
4.
Check that the diaphragm (
) is in the correct position by pulling
gently on it.
5.
Tighten the screws (
) crosswise applying a torque of approx. 1
Nm.
6.
Install the rubber tube piece (
) on the PVC tap on the membrane
disc (
). Do not use sharp tools for this.
7.
Install the guide sleeve (
). Screw in the screws (
) using a
little fitting grease (tightening torque approx. 1 Nm).
Diaphragm disc installed.
8.
Place the diaphragm disc (
) with the metal insert facing down-
wards in the lower housing (
) and ensure the diaphragm (
properly.
9.
Attach the upper housing (
). Applying slight pressure turn the
housing parts against each other to ensure that the diaphragm (
)
is well seated.
10.
Screw in the screws (
) with a little fitting grease. Tighten the
screws crosswise to a torque of approx. 2 Nm.
11.
Screw in the screw connection (
) by hand.
12.
Fit the O-ring (
) to the output and tighten the union nut (
) by
hand.
Vacuum part installed.
10.5.4 Testing the vacuum part
Testing for leaks
Precondition for action:
The vacuum part is installed.
The inlet valve has been successfully tested, thoroughly dried and
installed on the vacuum regulator.
The injector is ready for operation.
Fig. 10-4: Test for leaks
Perform the following working steps:
1.
Close the input of the vacuum regulator
2.
Connect the injector to the output of the vacuum regulator.
3.
Switch on the injector.
4.
After approx. ½ minute attach a tube to the safety valve connection.
5.
Immerse the open end in water.
The water must not rise in the tube.
Leak test concluded
Testing the functioning of the safety valve
Precondition for action:
The leak test is concluded.
Compressed air or nitrogen with a suitable pressure reducing valve
is available.
Fig. 10-5: Testing the safety valve
0 – 1 bar