DPX-IQ Operator’s Manual
Orthopedic Hip
G-7
The Line Profile
The Line Profile shows the information that follows:
•
Baseline
—The program calculates results from the area above the baseline (
➊
).
•
High Density Points
—The High Points on the profile show areas of high
density: dense bone (
➋
) and implant material (
➌
).
•
Edge Markers
—The edge markers (
➍
) are used to isolate the bone from the
implant. The high points contained inside the area between each pair of markers,
and above the baseline, are used to calculate results.
•
Soft Tissue Points
—The low points on the profile (
➎
) show areas of
soft tissue.
Figure G-7. Line Profile
Examine Edge Markers
Refer to the information that follows to make sure the edge marker positions are
correct (figure G-8):
•
Pelvis Edge
—If the pelvis is shown on the scan image, the markers are on each
edge of the pelvis bone points (
➊
and
➋
).
•
Left Femur Edge
—The left femur edge marker (
➌
) is on the last soft tissue
point before the bone points start up to the implant points.
•
Left Implant Edge
—The left implant edge maker (
➍
) is on the first
implant point.
•
Right Implant Edge
—The right implant edge marker (
➎
) is on the last
implant point.
•
Right Femur Edge
—The right femur edge marker (
➏
) is on the point
immediately after the bone points join the soft tissue points.
NOTE: Change edge markers only if they are obviously not correct. Let the
program calculate the edges for you.
➊
➋
➌
➍
➎
Summary of Contents for DPX-IQ
Page 6: ...vi DPX IQ Operator s Manual This page left blank intentionally...
Page 12: ...xii Table of Contents DPX IQ Operator s Manual...
Page 54: ...4 8 Default Settings DPX IQ Operator s Manual Figure 4 3 Analysis Results report 4 6d...
Page 56: ...4 10 Default Settings DPX IQ Operator s Manual This page left blank intentionally...
Page 68: ...A 12 AP Spine DPX IQ Operator s Manual This page left blank intentionally...
Page 79: ...DPX IQ Operator s Manual Femur B 11 Figure B 12 Position Neck ROI...
Page 80: ...B 12 Femur DPX IQ Operator s Manual This page left blank intentionally...
Page 100: ...D 10 Total Body DPX IQ Operator s Manual This page left blank intentionally...
Page 130: ...G 14 Orthopedic Hip DPX IQ Operator s Manual Figure G 15 Transverse Histogram screen...
Page 160: ...J 6 Precision and Accuracy DPX IQ Operator s Manual This page left blank intentionally...
Page 166: ...This page left blank intentionally...