FULLY INTEGRATED LED BASED  

UV DISINFECTION SYSTEM

Operation &  

Installation  

Instructions

GUV-4S / GUV-5S modules

Summary of Contents for Glacier UV LED GUV-4S

Page 1: ...FULLY INTEGRATED LED BASED UV DISINFECTION SYSTEM Operation Installation Instructions GUV 4S GUV 5S modules ...

Page 2: ...Quality Parameters 2 2 3 Location 2 3 0 Installation 3 4 0 Operation 4 4 1 Glacier GUV 4S Controller 4 4 2 Glacier GUV 5S Controller 5 4 3 Glacier GUV 5S Operational Screens 6 5 0 System Troubleshooting 7 8 6 0 System Specifications 9 7 0 Expansion Modules 10 8 0 Warranty 11 GUV 5S module GUV 4S module ...

Page 3: ...ts inside the GLACIER UV LED system Contact the manufacturer for service instructions if required 7 Do not use this system for any purpose other than what it was intended for Misuse of this system could potentially cause harm to the user or others Use of this system does not guarantee any particular amount of reduction of microorganisms 8 Your system is intended to be installed indoors and away fr...

Page 4: ...being treated UV Transmittance above 90 Hardness 1 gpm 17 mg L Iron Fe 0 05 ppm 0 05 mg L Manganese Mn 0 05 ppm 0 05 mg L Turbidity 0 1 NTU Tannins 0 1 ppm 0 1 mg L 2 3 Location For Point of Use POU systems install the unit just before the faucet The unit must be installed vertically for correct operation Ensure there s enough space to facilitate the plumbing fittings as well as any required pre t...

Page 5: ...VDC power adaptor into an appropriate AC electrical outlet To prevent risk of shock ensure that the AC outlet and power adaptor comply with local electrical safety codes 6 Plug in the 12VDC Adaptor to the Bottom Power Port The system should power on DO NOT OPERATE PRODUCT WITHOUT CONNECTING TO WATER SUPPLY 3 0 Water outlet tube Quick connect fitting for 3 8 OD tube Mounting screw Water inlet tube ...

Page 6: ... LEDs are disabled until the temperature drops below the Overtemperature Alarm Release Threshold During the cool down period the system no longer provides disinfection Refer to Operation and Specifications for details 4 1 GLACIER GUV 4S Controller Multi State Status LED indicating power flow LED life and overtemp fail alarms Automatic detection of water flow provides on demand disinfection low ene...

Page 7: ...On Sequence On Start up the system will run through a diagnostic check and initialize any optional module that may be attached to the system Optional Modules Solenoid WiFi Remote Alarm Sherpa SHERPA OR Solenoid WiFi Remote Alarm A final module screen is displayed showing which specific modules were initialized and then proceed to normal system operation The total length of the start up procedure i...

Page 8: ...the UV LEDs This displayed as LED Life Hours remaining Each Systems UV LEDs have up to 5000 Hours Lifetime TEMPERATURE SCREEN Displays the temperature of the UV LEDs CONTACT SCREEN Displays company information QR CODE A QR code Quick Response code is a matrix barcode first designed for the automotive industry LUMINOR uses the QR code to store a link to a specific page on our website INFO SCREEN Di...

Page 9: ...e back or on the Info Screen NO Constant Beeps Although the UV LEDs are powered and visibly illuminated due to the LEDs age its UV output is no longer sufficient for proper disinfection Replace unit with the part indicated on the silver label on the back or on the Info Screen NO Constant Beeps The UV LEDs are overheating the system enters a cool down period disabling the UV LEDs until the temperat...

Page 10: ...ling power to the system Replace any module that is not detected when plugged directly into the controller YES 2 beeps every 30 seconds The module indicated is no longer communicating to with the system YES 2 beeps every 30 seconds The UV LED remaining hours is less than 8 but has yet to expire User needs to start consider replacing unit soon The part number is indicated on the silver label on the...

Page 11: ...t style quick connect fittings Input Voltage 12VDC Input Power 15W 16W Electrical Connection Standard 12VDC barrel connector 2 1 x 5 5 mm UV Intensity Monitoring NA Coming soon LED Status Monitoring YES LED Life Monitoring YES LED Life 5 000 hours Solenoid Output NA YES optional module MOD SOL sold separately Dry Contacts solenoid ready NA YES optional module MOD RAM sold separately 4 20 mA Output...

Page 12: ...re alarm The solenoid closes during an alarm that cause Unsafe Water Conditions Also note that in cases where emergency use of untreated water is required the controller can be placed into a manual override mode allowing for the flow of water in an alarm condition Maximum voltage and load on the MOD SOL is 240V 2A Order PN MOD SOL WiFi module and accompanying IoT application Allows you to connect ...

Page 13: ...anty LUMINOR will not be liable for damage or wear to products caused by abnormal operating conditions accident abuse misuse unauthorized alteration or repair or if the product was not installed in accordance with LUMINOR s printed installation and operating instructions LIMITATIONS OF WARRANTY This warranty does not apply to any of the following Water Quality Parameters lie outside of the followi...

Page 14: ......

Page 15: ......

Page 16: ...80 Southgate Drive Unit 4 Guelph Ontario Canada N1G 4P5 P 519 837 3800 TF 855 837 3801 E info luminoruv com www luminoruv com Rev 05 2020 EPA Establishment 088776 CAN 001 ...