![LumaSense Impac IGA 5 User Manual Download Page 20](http://html1.mh-extra.com/html/lumasense/impac-iga-5/impac-iga-5_user-manual_1947571020.webp)
INFRATHERM
pyrometers IS 5 · IS 5-LO · IGA 5 · IGA 5-LO
49
•
The devices IS 5-C and IGA 5-C have an additional PID-controller. The functions of this controller can be
activated with a click on the “PID” icon (
).
The open / save button enable to store and recall own pyrometer configurations.
„1 meas.“ shows the current measuring temperature in the pyrometer parameters window for
approx. 1 second.
A click on the “Test” icon opens a window that
allows the direct communication with the py-
rometer via the interface commands (see
section
Data format UPP
®
).
After entering an interface command (00 is the
adjusted address ex works, “ms“ is the com-
mand “reading temperature value“) and a click
on “Send“ the displayed window is opened:
This window already shows the answer of the
pyrometer in
1
/
10
°. The actual temperature
reading is 325.7°C.
“Len“ indicates the length of the answered data string, incl. Carriage Return (Chr(13)).
In the lower part of the window the connection with the preset baud rate can be checked. Here the command
was send 500 times with 19200 baud. It has taken 4.56 seconds and no transmission errors has occurred.
10.8
Configuration of the display on the TV screen
The video screen with the available display options can be configured here
(only models with built-in color video module).
•
Text: input of a text consisting of max.
12 figures, e.g. machine07
•
Serial No.: the serial number of the instrument
can be displayed automatically instead of the
text
•
Setting of the actual time and date
•
Brightness: Selection of brightness adjustment
of the video image between automatic or man-
ual mode
The video image is used for alignment of the
pyrometer onto the measuring object and shows
the following:
•
measuring object and ambient area
•
spot mark
•
current temperature reading
•
adjusted emissivity of the pyrometer
•
actual time and date
•
text or serial number
Note:
The display of the time (12 or 24 h mode) is
depending on the adjustment of the temperature
scale (°C / °F) of the pyrometer:
24 hour display is combined with °C, 12 hour
display with “am” and “pm” is combined with °F.