LumaSense Impac IGA 5 User Manual Download Page 10

 

INFRATHERM

 pyrometers IS 5 · IS 5-LO · IGA 5 · IGA 5-LO 

 

 

39

IS 5; IGA 5 
IS 5-LO; IGA 5-LO

 

230V ~

24 V DC

 

°C

 

yellow

 

LED digital 
display 

 

controller

power supply

green

 

brown

 

white

 

printer

4.5 

Connection schematic for analysing devices  

Additional analyzing instruments, for 
example a LED digital display instru-
ment only needs to be connected to a 
power supply and the analog outputs 
from the pyrometer (exception: the 
digital display DA 6000 also can be 
connected with its serial interface, the 
digital display DA 6000-N has to be connected with its serial interface). 
Another Instruments like a controller or printer can be connected to the 
display in series as shown above (total load of resistance max. 500 
Ohm). 
 
 

5 Mechanical 

installation 

 

5.1 Accessories 

(option) 

Numerous accessories guarantee easy installation of the pyrometers. The following overview shows a selec-
tion of suitable accessories. You can find the entire accessory program with all reference numbers on section 

15.2

,

 Reference numbers

 
Mounting (for IS 5; IGA 5): 

For easy mounting and aligning the pyrometer to the 
measured object an adjustable 

mounting angle

 is avail-

able. 

Mounting angle 

 
Cooling (for IS 5; IGA 5):

 

The completely covered water cooling jacket made from 
stainless steel protects the pyrometer if exposed to a hot 
environment. It is designed for ambient temperatures up 
to 180°C. 

Water cooling jacket with 
integrated air purge

 

 

Miscellaneous (for IS 5; IGA 5): 

The 

air purge

 protects the lens from contamination with 

dust and moisture. It has to be supplied with dry and oil-
free pressurized air and generates an air stream shaped 
like a cone. 
The pyrometer can be easily fixed on a vacuum chamber 
with the KF 16 vacuum support with sighting window.

 

 

 

Air purge

 

 

 

 

Vacuum support

 

 
The 

scanning mirror unit SCA 5

 

(for IS 5; IGA 5) 

enables the 

measured object to be scanned over a certain range. The measuring 
beam of the pyrometer moves straight in one line across the object 
and collect temperature data of this line. This is useful in combina-
tion with the maximum value storage (peak picker) to measure ob-
jects which move out of the target area. The scanning angle of the 
mirror is 0 ... 12°, the scanning frequency 0 ... 5 Hz. Both values are 
easily adjustable at the instrument. 

 

 

 
The 

mounting tube

 

(for  IS  5;  IGA  5) 

is used for fixing of the py-

rometer at machines and protects the instrument. It is equipped with 
an air connector to cool the pyrometer with compressed air and to 
purge the optics to keep it clean.  
The 

flange tube

 is similar to the mounting tube, it will be delivered 

with an additional flange for fixing. This flange has to be welded in 
the required position. 

Summary of Contents for Impac IGA 5

Page 1: ...IS 5 IGA 5 IS 5 LO IGA 5 LO INFRATHERM Pyrometer IMPAC Spezialist f r ber hrungslose Temperaturmessung IMPAC Specialist in non contact thermometry Betriebsanleitung User Manual...

Page 2: ...S485 38 4 5 Connection schematic for analysing devices 39 5 Mechanical installation 39 5 1 Accessories option 39 5 2 Connection overview IS 5 LO IGA 5 LO 40 6 Instrument alignment 41 6 1 View finder 4...

Page 3: ...48 10 8 Configuration of the display on the TV screen 49 10 8 1 The PID controller 50 10 8 2 Self tuning algorithm 51 10 9 Measurement color bar 51 10 10 Measurement online trend 52 10 11 Output listi...

Page 4: ...otify any person or persons of such revisions or changes All series 5 Instruments from IMPAC Infrared GmbH have a warranty of two years from the invoice date This warranty covers manufacturing defects...

Page 5: ...isolated from each other Parameters Adjustable on the pyrometer Emissivity exposure time t90 analog output 0 20 or 4 20 mA online offline switch for adjustments at pyrometer or PC Additionally adjusta...

Page 6: ...levels controlled by the measuring temperature Field of view approx 10 x 14 of focused distance Date time real time clock with at least 3 days spare run Video output plug separate round plug at the p...

Page 7: ...if the person already used similar instruments or was already trained by the manu facturer The pyrometer has only to be used for the purpose described in the manual It is recommended to use only acces...

Page 8: ...e via external switch J pink see 4 1 1 external clearing of max value storage or hold function Only pyrometers with PID controller controller activated set controller output to 0 emergency stop G red...

Page 9: ...r fast measuring equipment So it is recommended to use the IMPAC converter order no 3 852 430 4 3 Connection via interface RS232 The transmission rate in baud of the serial interface is dependent on t...

Page 10: ...made from stainless steel protects the pyrometer if exposed to a hot environment It is designed for ambient temperatures up to 180 C Water cooling jacket with integrated air purge Miscellaneous for IS...

Page 11: ...way to align the pyrometer to the measured object The clamping screws of the ball and socket mounting enable an easy and fast adjustment of the pyrometer in all directions Mounting angle Ball and sock...

Page 12: ...Attention Do not look directly into the laser beam Laser class 2 according to IEC 60825 1 3 4 Note The laser warning signs on the pyrometer should be easily viewable at all times even after it has bee...

Page 13: ...2 15 7 2 Differing to the focused measuring distance all instrument types The optical profiles show that the spot size M varies with the distance from the pyrometer to the measuring object Spot sizes...

Page 14: ...out bending or twisting it Caution Disconnect the cable before opening the cover When reassembling the cover insert it carefully into the guide pin and then fasten it with the screws After that the co...

Page 15: ...r a correct measurement it is necessary to adjust the emissivity This emissivity is the relationship between the emission of an real object and the emission of a black body radiation source this is an...

Page 16: ...ttings The digital PC interface enables the exchange of data with a PC either by using the InfraWin software sup plied or an user adapted communication software see command table in section 14 Data fo...

Page 17: ...stab lished after each impulse from the reset signal double storage when entering the reset intervals via push buttons or PC interface the double storage is automatically selected This mode utilizes t...

Page 18: ...n of the individual icons in the program s help menu Click on the F1 button after InfraWin has been loaded or click on the in the menu bar The following descriptions refer to program version 4 0 The l...

Page 19: ...of devices InfraWin changes to the display of 1 or 2 devices If 2 devices are selected always 2 windows are displayed for settings or evaluation 10 7 Basic settings Under pyrometer parameters all pre...

Page 20: ...indow the connection with the preset baud rate can be checked Here the command was send 500 times with 19200 baud It has taken 4 56 seconds and no transmission errors has occurred 10 8 Configuration o...

Page 21: ...sired temperature a corresponding signal which changes over the time is transmitted to the analog output until the deviation reaches zero The integral time can be set in two different formats min s fo...

Page 22: ...red temperature Problems are generally posed by directly controlled systems where there are high contrasts between the cooling time constant and the heating up time constant For example in many cases...

Page 23: ...cally saved as standard i12 Should you need to edit the data later you need to save the file as another i12 file because old values are over written when a new measurement is taken Files from older pr...

Page 24: ...er file the previous file will be overwritten and replaced by the standard i12 file 10 14 PC sampling rate time interval between two measurements This function sets a time interval After each interval...

Page 25: ...lower correct emissivity corresponding to the material see 8 5 Lens contaminated Clean lens carefully see 12 2 Temperature indication too high Emissivity set too low Set lower correct emissivity corre...

Page 26: ...repeated AAmsXXX XXX 000 999 auto repeat of the ms command Reading basic temperature range AAmb Output XXXXYYYY 2 x 4 hexdigit for MBA and MBE C or F Reading sub range AAme Output XXXXYYYY 2 x 4 hexdi...

Page 27: ...rtional control Reset time Integral time AATiXX XX AATiXX XX or XX XX Ti time in format min sec XX XX is equivalent to 00 01 99 99 calculation is every second Ti time in Format XX XX sec is equivalent...

Page 28: ...r RS485 IS 5 MB 20 600 2000 C 3 857 150 3 857 170 3 857 160 3 857 180 IS 5 MB 25 800 2500 C 3 857 200 3 857 220 3 857 210 3 857 230 IS 5 MB 30 1000 3000 C 3 857 250 3 857 270 3 857 260 3 857 280 IGA 5...

Page 29: ...justable for optical head I 3 820 330 Connection cable 5 m long straight pyrometer connector 3 820 500 Connection cable 10 m long straight pyrometer connector 3 820 510 Connection cable 15 m long stra...

Page 30: ...ent 41 Instrument settings 43 Interface commands 49 Interface connection 38 Internal temperature of the pyrometer 46 L Laser targeting light 36 41 M Maintenance 54 Maximum value storage 37 45 Measurem...

Page 31: ...IMPAC Infrared GmbH Temperaturmessger te Kleyerstra e 90 D 60326 Frankfurt Main Tel 49 0 69 9 73 73 0 Fax 49 0 69 9 73 73 167 Internet www impacinfrared com E Mail info impacinfrared com 3 869 207 a...

Reviews: