background image

 

IN 5/9 plus Manual 

Controls and Installation 

  14 

commonly used adapters are too slow for fast measuring equipment, so it is recommended to 

use the RS485 

 USB adapter cable (ref. no. 3 826 750). 

With a slow RS485 connection it is also possible to set a wait time at the pyrometer, which 
delays the response of a command to the pyrometer (see 

5.2.6 Wait time t

w

). 

3.1.3

 

Connecting to RS232 interface 

The transmission rate (in baud) of the serial 
interface is dependent on the length of the cable. 

Values between 1200 and 19200 Bd may be set. 
The baud rate must be reduced by 50% when the 
transmission distance is doubled (see also 

5.2.3 

kBaud (baud rate)

). 

Typical cable length for RS232 at 19200 Bd is 7 m. 

3.1.4

 

Connecting to RS485 interface 

Half-duplex mode: 
A1 and A2 as well as B1 and B2 are 

bridged in the 12-pin round 
connector of the connecting cable, 

to prevent reflections due to long 
stubs. It also safeguards against the 

interruption of the RS485 Bus 
system should a connecting plug be 

pulled out. The master labels mark 
the connections on the RS485 

converter. The transmission rate of 
the serial interface in Baud (Bd) is 

dependent on the length of the 
cable. Values between 1200 and 

19200 Bd may be set.  
The baud rate is reduced by 50% 
when the transmission distance is doubled (see 

5.2.3 Baud Rate (kBaud)

).

 

Typical cable length 

for 19200 Bd is 2 km. 

3.1.5

 

Connection of additional analyzing devices 

Additional analyzing instruments (such as LED digital display instruments) only need to be 
connected to a power supply and the analog outputs from the pyrometer. Another Instrument, 

such as a controller or printer, can be connected to the display in series as shown below (total 
load of resistance max. 500 Ohm). 

 

yellow

 

LED digital display

 

Controller

 

Power supply

 

green

 

brown

 

white

 

230 V ~

 

24 V DC

 

°C

 

Writer

 

Converter 

 

 

K

 

 

 

A

 

 

 

B

 

 

 

C

 

 

 

D

 

 

 

J

 

 

 

G

 

 

 

F

 

 

 

E

 

 

 

H

 

 

 

M

 

 

 

L

 

 

 

1

 

 

 

2

 

 

 

3

 

 

 

4

 

 

 

5

 

 

 

6

 

 

 

7

 

 

 

8

 

 

 

9

 

 

 

 DGND (red)

 

 RxD (black)

 

 TxD (violet)

 

Pyrometer’s side 

(soldering side)

 

cable

 

PC’s side 

(soldering side)

 

 

Terminator 120 Ohm

 

A

 

B

 

S

 

DG

ND

 

B

1

 

A

2

 

B

2

 

A

1

 

DG

ND

 

B

1

 

A

2

 

B

2

 

A

1

 

DG

ND

 

A

2

 

B

2

 

A

1

 

B

1

 

Master

 

Pyrometer 32 

e.g. address 31

 

Pyrometer 2 

e.g. address 01

 

Pyrometer 1 

e.g. address 00

 

Summary of Contents for IMPAC IN 5 plus

Page 1: ...MANUAL IMPAC Pyrometer IN 5 9 plus...

Page 2: ...spective holders Service Centers LumaSense Technologies Inc North America Sales Service Santa Clara CA USA Ph 1 800 631 0176 Ph 1 408 727 1600 Fax 1 408 727 1677 LumaSense Technologies GmbH Other Than...

Page 3: ...ons 10 2 5 Physical user interface 11 2 6 Accessories Optional 11 2 6 1 Mounting 11 2 6 2 Water Cooling Jacket 11 2 6 3 Miscellaneous 12 2 6 4 Display 12 3 Controls and Installation 13 3 1 Electrical...

Page 4: ...ar 20 5 2 2 Subrange 21 5 2 3 Baud rate kBaud 21 5 2 4 Address 22 5 2 5 Ambient temperature compensation 22 5 2 6 Wait time tw 23 5 2 7 Internal temperature 23 5 3 Avoiding Reading Errors 23 6 Softwar...

Page 5: ...e in German Messbereich 1 1 2 Terminology The terminology used in this manual corresponds to the VDI VDE directives 3511 Part 4 1 2 Safety This manual provides important information on safely installi...

Page 6: ...ocument and to make changes from time to time in the content hereof without obligation to notify any person or persons of such revisions or changes All instruments from LumaSense Technologies have a r...

Page 7: ...elephone If Technical Support concludes that the instrument must be returned to LumaSense Technologies for repair they will issue a Return Material Authorization RMA number Return the instrument upon...

Page 8: ...he instrument can be damaged or destroyed To transport or store the instrument please use the original box or a box padded with sufficient shock absorbing material For storage in humid areas or shipme...

Page 9: ...mb 0 15 or 30 63 C Whichever value is greater The instrument must be at a constant ambient temperature for a minimum of 60 minutes and must be connected to the power supply Repeatability 1 t90 1 s Tam...

Page 10: ...Supply 24 V DC 18 30V DC nominal ripple must be less than 0 5 V Power Consumption Max 70 mA Isolation Power supply analog outputs and digital interfaces are electrically isolated from each other Ambi...

Page 11: ...g overview shows a selection of suitable accessories You can find the entire accessory list with all reference numbers in Section 10 2 Reference numbers accessories 2 6 1 Mounting For easy mounting an...

Page 12: ...enerates an air stream shaped like a cone The pyrometer can be easily mounted on a vacuum chamber with the KF 16 vacuum support with sighting window Air purge units Vacuum support 2 6 4 Display In add...

Page 13: ...output B yellow Ioutp analog output H gray external switch for targeting light bridge to K J pink external clearing of maximum value storage bridge to K G red DGND Ground for interface F black RxD RS2...

Page 14: ...out The master labels mark the connections on the RS485 converter The transmission rate of the serial interface in Baud Bd is dependent on the length of the cable Values between 1200 and 19200 Bd may...

Page 15: ...ontact see 3 1 1 Pin assignment for the connector on the back side of the pyrometer or via PC and the software InfraWin After two minutes the laser targeting light switches off automatically Caution D...

Page 16: ...12 7 350 23 7 4 270 4 5 500 19 7 750 36 9 5 400 6 4 750 25 1000 38 2 6 620 10 1000 25 2 1500 45 2 Effective aperture D for all optics is 14 8 mm For each optic some example values for measuring distan...

Page 17: ...emissivity exposure time analog output maximum value storage minimum value storage reading of the instrument s internal temperature setting of an address for bus control with RS485 interface setting...

Page 18: ...de ONL only the serial interface can be used to set the values for the parameters and the instrument s switch settings are ignored The device is initialized in online mode i e it uses the last values...

Page 19: ...etting examples at the instrument Emi 100 Emi 85 If the settings directly at the pyrometer should be used the pyrometer must be switched into offline mode OFFL see 4 2 To adjust the emissivity via sof...

Page 20: ...nd stored The minimum value storage saves the lowest measurement taken during a reading e g practical when monitoring cooled goods such as frozen foods As such it may be beneficial to periodically cle...

Page 21: ...has been exceeded The minimum value function operates in the opposite manner The minimum value stored will be deleted when the upper limit of the set range is below it by 1 at least 2 C If an upper l...

Page 22: ...to be measured is in an oven where the temperature is higher than the ambient temperature a portion of the radiation corresponding to the rate of diffuse reflection will be reflected by the measured o...

Page 23: ...3 Optics 2 A source of radiation behind or around the measuring object can influence the result If the object is transparent or partly transparent another material behind the object could transmit it...

Page 24: ...IN 5 9 plus Manual Parameter Descriptions Settings 24 To ensure consistent document formatting this page was intentionally left blank...

Page 25: ...c com 6 1 Connecting the pyrometer to a PC The program InfraWin can operate up to two devices Two devices using RS485 may be operated simultaneously by the same interface if two different addresses ha...

Page 26: ...IN 5 9 plus Manual Software InfraWin 26 To ensure consistent document formatting this page was intentionally left blank...

Page 27: ...aned with standard lens tissue dampened with a commercially available glasses or camera lens cleaning solution Use a soft blower brush available in camera stores to remove any grit on the window befor...

Page 28: ...IN 5 9 plus Manual Maintenance 28 To ensure consistent document formatting this page was intentionally left blank...

Page 29: ...ment is influenced by reflections of hotter machine parts Try to avoid the influence of the interfering radiation or change measuring position Measuring errors Indicated temperature is decreasing duri...

Page 30: ...IN 5 9 plus Manual Troubleshooting 30 To ensure consistent document formatting this page was intentionally left blank...

Page 31: ...letters answers with the respective settings only at setting commands not at enquiry commands Example Entry 00em CR The emissivity setting of the device with the address 00 is returned Answer 0970 CR...

Page 32: ...Clear time max storage tclear Digit 5 0 1 Analog output Digit 6 and 7 00 98 Temperature Digit 8 and 9 00 31 Address Digit 10 0 4 Baud rate Digit 11 always 0 Error status AAfs Output 1 Byte hex Bit 0...

Page 33: ...aser targeting light push button 3 820 740 Connection cable 5 m straight connector temperature resistant up to 200 C 3 852 290 Power supply NG DC for DIN rail mounting 100 to 240 V AC 24 V DC 1 A 3 85...

Page 34: ...jacket heavy design with integrated air purge unit 3 837 350 Heavy water cooling jacket with protection window 3 837 370 Water cooling jacket lightweight design with integrated air purge unit 3 837 40...

Page 35: ...ectrical connection 6 Electrical Installation 13 Emissivity 19 Exposure time 19 F Factory setting 18 G General Information 5 I Installation Electrical 13 Instrument settings 17 Instrument settings 18...

Page 36: ...criptions 19 Shield 13 Sighting 15 Software InfraWin 25 Spot size 16 Storage 8 Subrange 21 Support 7 Switches at the Instrument 17 T Technical Data 9 Transport 8 Troubleshooting 29 U Universal Pyromet...

Reviews: